Z-0913-2015Class II

The Symbia Intevo Excel is a non-diagnostic SPECT ICT system with CT support for only attenuation correction...

Siemens Medical Solutions USA, Inc. / initiated 2014-08-12 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc. began a recall of The Symbia Intevo Excel is a non-diagnostic SPECT ICT system with CT support for only attenuation correction and anatomical localization. The Symbia Intevo... on . FDA classified it as Class II and gives the reason as "Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0913-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
The Symbia Intevo Excel is a non-diagnostic SPECT ICT system with CT support for only attenuation correction and anatomical localization. The Symbia Intevo Series are xSPECT systems. These are SPECT and T systems integrated through xSPECT technology during image registration and reconstruction. This xSPECT integration enables the use of extra-modal information for high image quality and quantification. The Siemens Symbia series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures. SPECT: To detect or image the distribution of radionuclides in the body or organ, using the following techniques: planar imaging, whole body imaging, tomographic imaging for isotopes with energies up to 5S8keV. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. SPECT+CT: Perform CT scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy. Software: The MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies taken from SPECT, PET, CT and other imaging modalities.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent Customer Safety Advisory Notice letters to all customers of record. The letters instructed customers to: 1) continue to use their system(s); 2) place the safety advisory in the system's instructions for use; and, 3) to anticipate that their local service engineer will contact them to perform an inspection and schedule any necessary repairs by December 2014. Customers were also instructed to share the letter with their support staff and, if the equipment was sold or transferred, to forward the safety notice to the new owner of the equipment. Customers with questions can contact their local service representative or Siemens Medical Solutions at 800-888-7436.
Quantity in commerce
6 systems
Distribution
Worldwide Distribution - USA (naitonwide) and Internationally to Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Latvia, Lithuania, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, and Vietnam.
Product codes and lots
1) Symbia Intevo 6: Product Number: 10764803; Serial Numbers: 1001 ; 1002 ;  2) Symbia Intevo 16; Product Number: 10764804; Serial Numbers: 1001 ; 1002 ;  3) Symbia Intevo Excel; Product Number: 10764801; Serial Numbers: 1001 ; 1002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0913-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0911-2015Class IITerminatedVoluntary: Firm initiated
Z-0912-2015Class IITerminatedVoluntary: Firm initiated
Z-0913-2015Class IITerminatedVoluntary: Firm initiated