SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids
Baxter Healthcare Corp. began a recall of SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids on . FDA classified it as Class II and gives the reason as "One Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2738-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- SPECTRUM Pump, Model No. 35700BAX. Intended to be used for the controlled administration of intravenous fluids.
- Product code
- FRN Pump, Infusion
- Reason for recall
- One Service Technician may not have correctly serviced specific Sigma Spectrum Infusion Pumps according to established procedures during the time period of 5/5/2014 through 6/3/2014.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- Initial notification was initiated via telephone call by the Baxter Medina Service Center on 8/28/14 to all affected Sigma SPECTRUM Infusion Pump consignees. URGENT DEVICE CORRECTION Letters (dated 9/03/2014) were sent to the consignees via USPS first class mail on 9/03/14 formally informing them of the recall. The letters identified the affected product, the description of the issue, the hazard involved, as well as, the actions to be taken by customer. Customers are being instructed to immediately remove the pump from use and return the pump to Baxter for inspection. The firm will provide replacements. Customers are to contact Baxter Healthcare Medina at 800-356-3454 for technical questions regarding the letter.
- Quantity in commerce
- USA: 56 units, Canada: 2 units
- Distribution
- Worldwide Distribution -- USA and Canada.
- Product codes and lots
- Software Versions 5.02.06, 6.02.06, and 6.02.11; Affected Serial Numbers: 712090, 723687, 723842, 724966, 725820, 735977, 751130, 752124, 755174, 768538, 771990, 774743, 781406, 783736, 784698, 794466, 805797, 808901, 809113, 812462, 815932, 842120, 857926, 858672, 862085, 863721, 865630, 873459, 889597, 890260, 903748, 912619, 920574, 920589, 956346, 957394, 965402, 966684, 976931, 977550, 978361, 983979, 984066, 984129, 984475, 985946, 987538, 993445, 995291, 996014, 996389, 1013037, 1004377, 1014565, 1014962, 967887, 950671, and 938428.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2738-2014 | Class II | Terminated | Voluntary: Firm initiated |