Z-2665-2014Class II

Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network...

Siemens Healthcare Diagnostics Inc. / initiated 2014-08-19 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and multi-system data manager for... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management System, software version 14. The issue may cause the software to stop executing commands, including uploading...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2665-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens CentraLink Data Management System, software version 14. The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab.
Product code
N/A
Reason for recall
Siemens Healthcare Diagnostics has identified a task scheduler issue with the CentraLink" Data Management System, software version 14. The issue may cause the software to stop executing commands, including uploading validated results to the LIS. The issue is related to an internal software timer that overflows after 24 days. Automated rules triggered by task schedulers or by actions in the CentraLink user interface may also be affected. As each customer configuration is unique, in addition to the delay in uploading of validated results, the specific consequences of discontinued command execution will vary.
Root cause tag
software
FDA root cause
Software design
Action
An Urgent Medical Device Correction (UMDC #10819364) letter was issued on August 19, 2014 to all affected customers. The letter instructed customers as follows: Siemens Healthcare Diagnostics is conducting a field correction for the CentraLink" Data Management System. To prevent the issue from occurring, log out of user sessions and restart the CentraLink services and translators as soon as it is convenient, then periodically restart going forward. Any restart frequency that occurs more often than every 24 days is acceptable and will prevent the issue from occurring. Refer to the Additional Information section for detailed instructions.
Quantity in commerce
858
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Egypt, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Lebanon, Lithuania, Luxemburg, Malaysia, Mexico, Netherland, New Zealand Norway, Pakistan, Portugal, Peru, Qatar, Russia, Saudi Arabia, Singapore Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand United Arab Emirates, United Kingdom, Uruguay, Zambia.
Product codes and lots
Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2665-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2665-2014Class IITerminatedVoluntary: Firm initiated