510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2688-2014Class II

RENASYS EZ Negative Pressure Wound Therapy (NPWT) System (RENASYS EZ) for patients who would benefit from a...

Smith & Nephew, Inc. / initiated 2014-08-13 / Terminated

Smith & Nephew, Inc. began a recall of RENASYS EZ Negative Pressure Wound Therapy (NPWT) System (RENASYS EZ) for patients who would benefit from a suction device particulary as the device may... on . FDA classified it as Class II and gives the reason as "Modification were made to the RENASYS EZ NPWT Systems without a 510(k) or PreMarket Approval notification". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2688-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
RENASYS EZ Negative Pressure Wound Therapy (NPWT) System (RENASYS EZ) for patients who would benefit from a suction device particulary as the device may promote wound healing.
Product code
OMP Negative Pressure Wound Therapy Powered Suction Pump
Reason for recall
Modification were made to the RENASYS EZ NPWT Systems without a 510(k) or PreMarket Approval notification.
FDA root cause
No Marketing Application
Action
All RENASYS EZ Products will be removed from the field and customers will be transitioned to other product. On August 13, 2014 the firm sent an "Urgent- Medical Device Removal RENASYS EZ" notification via FedEx overnight asking that the use of the RENASYS EZ should be immediately discontinued and the device should be returned to Smith & Nephew. If you have questions, please reach out to your Customer Care representative or call 1-800-876-1261 Monday through Friday, 8:00 AM to 6:00 PM, Eastern Time.
Quantity in commerce
2,142 units
Distribution
Worldwide Distribution- US (nationwide) including MA, MN, NY and Puerto Rico; Canada, Mexico, Dubai, Australia, Chile, Colombia, Germany, France, England, Spain, Italy, Norway, Hong Kong, Shanghai, New Zealand, South Africa, Singapore, Japan, Thailand and the Caribbean Islands.
Product codes and lots
Model #'s 66800059, all lots since product launched 01/16/2009

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2688-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2688-2014Class IITerminatedVoluntary: Firm initiated