Z-2659-2014Class II
ZOLL Pro-padz Sterile Adult Multi-Function Electrodes ((singles) Part number: 8900-4013 These MFE Electrodes...
ZOLL Medical Corporation / initiated 2014-08-26 / Terminated / root cause: manufacturing process
Bio-Detek, Inc. began a recall of ZOLL Pro-padz Sterile Adult Multi-Function Electrodes ((singles) Part number: 8900-4013 These MFE Electrodes are used in connection with ZOLL Defibrillators... on . FDA classified it as Class II and gives the reason as "Sterility of device is not assured". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2659-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZOLL Medical Corporation
- Product
- ZOLL Pro-padz Sterile Adult Multi-Function Electrodes ((singles) Part number: 8900-4013 These MFE Electrodes are used in connection with ZOLL Defibrillators for Adult patients. These disposable single use MFE's are processed through Gamma Radiation to a SAL of 1 o.e per International/ National Standards, EN ISO 11137- 2:2012 and ANSI/AAMI/ISO 11137-2:2013, Method 1, for use in a sterile environment by trained medical professionals. The indications for use include defibrillation, cardioversion, external non-invasive pacing and monitoring. These MFE electrodes have a 12 month shelf life from time of manufacture.
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Sterility of device is not assured
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- BioDetek sent an Urgent Device Correction letter dated August 26, 2014, to all affected customers via UPS overnight delivery. There has also been notifications addressed to the 4 International accounts on August 27, 2014. Customers were requested to return the suspect product(s) in their possession and complete the Device Correction Response Return Form provided to each customer. Customer inventory will be replaced as soon as replacement electrodes are processed. Customers with questions can contact the technical support numbers at 1 (800) 348-9011 or +1 (978) 421-9460 24/7. For questions regarding this recall call 401-729-1400.
- Quantity in commerce
- 12 each
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Canada, and Germany.
- Product codes and lots
- Lot Number: 2314
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2658-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2659-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2660-2014 | Class II | Terminated | Voluntary: Firm initiated |