Z-0064-2015Class III

Omni Micro-electrode/reference electrode for cobas b221 analyzer. The Roche Diagnostics Omni S Analyzer is a...

Roche Diagnostics GmbH / initiated 2014-09-12 / Terminated / root cause: labeling and UDI

Roche Diagnostics Operations, Inc. began a recall of Omni Micro-electrode/reference electrode for cobas b221 analyzer. The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to... on . FDA classified it as Class III and gives the reason as "The default references for normal values are inconsistent between cobas b 221 and other blood gas instruments. In addition, the reference ranges are different when comparing with corresponding ranges listed in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0064-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Omni Micro-electrode/reference electrode for cobas b221 analyzer. The Roche Diagnostics Omni S Analyzer is a fully automated critical care analyzer intended to be used for the measurement of pH, PCO2, PO2, sodium, potassium, ionized calcium, chloride, hematocrit, glucose, lactate, urea/BUN, total hemoglobin, oxygen saturation, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin in samples of whole blood, serum, plasma, and aqueous solutions as appropriate.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
The default references for normal values are inconsistent between cobas b 221 and other blood gas instruments. In addition, the reference ranges are different when comparing with corresponding ranges listed in the Instructions for Use for the analyzer and the reference literature source.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
The Urgent Medical Device Correction will be sent via UPS Ground (receipt required) to the consignee list. The Urgent Medical Device Correction will be sent via UPS Ground (receipt required) to the consignee list. Consignees are instructed as follows: " Check the current set of reference values on your cobas b 221 blood gas system and, if appropriate, change the reference ranges accordingly. " If you have not established your own reference ranges, please follow the instructions in the Workaround section of the UMDC or contact the Roche Support Network Customer Support Center.
Quantity in commerce
492 devicees
Distribution
Nationwide Distribution including Puerto Rico and Guam, and the states of AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, ME, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA and WI.
Product codes and lots
Model/Catalog/Part Number: 1. 03337103001 2. 03337111001 3. 03337154001 4. 03337138001 5. 03337120001 6. 03337146001 7. 03337111692 8. 03337138692 9. 03337154692

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0064-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0064-2015Class IIITerminatedVoluntary: Firm initiated