Z-2711-2014Class II

The Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are Computed Tomography X-ray systems...

Philips North America LLC / initiated 2014-09-11 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of The Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are Computed Tomography X-ray systems intended to produce cross-sectional images of the body... on . FDA classified it as Class II and gives the reason as "Ingenuity CT system can mislabel the exposure that results in an image in the incorrect order that it was taken by the user". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2711-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
The Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Ingenuity CT system can mislabel the exposure that results in an image in the incorrect order that it was taken by the user.
Root cause tag
software
FDA root cause
Software design
Action
Philips sent an Urgent - Field Safety Notice Medical Device Correction letter dated September 5, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please retain a copy with the equipment Instruction for Use. Additionally, a Field Service Engineer will contact each site to schedule a time to install the software update. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America, contact the Customer Care Solutions Center (1.800.722-9377 and follow the prompts).
Quantity in commerce
79
Distribution
Worldwide Distribution - US including the states of : AZ, CA, CT, FL, GA,MA, MS,OH, PA, TX, and WY., and the countries of : Australia, Denmark, Finland, France, Germany, India, Ireland, Japan, Malaysia, Netherlands, New Zealand, Norway, Russia, Slovakia, South Africa, South Korea, Sweden, Switzerland and United Kingdom.
Product codes and lots
Ingenuity CT: 728326 300005,300010,300012,300016,300017,300019,300020,300021,300028,300029, 300031,300032,300033,300035,300037,300039,300043,300047,300050,300051, 300052,300054,300055,300057,300060,300061,300062,300063,300072,300074, 300082,300086,300087,300089,300094,300095,300100,300101,300102,300106, 300108,300110,300112,300113,300129,300135,300136,300140,300143  Ingenuity Core: 728321 310001,310002,310009,310055,310059,310064,310069,310070,310087,333015  Ingenuity Core128: 728323 300137,320003,320005,320006,320018,320024,320029,320030,320033,320037, 320078,320092,320099,320103,320114,336011,336012,336013,336016,336018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2711-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2711-2014Class IITerminatedVoluntary: Firm initiated