Z-0201-2015Class II

Integra Licox Pt02 Monitor, Rx Only, Manufacturer: Integra LifeSciences (Ireland) Limited, IDA Business and...

Integra LifeSciences Corporation / initiated 2014-08-18 / Terminated / root cause: design

Integra LifeSciences Corp. began a recall of Integra Licox Pt02 Monitor, Rx Only, Manufacturer: Integra LifeSciences (Ireland) Limited, IDA Business and Technology Park, Sragh, Tullamore, County Offaly... on . FDA classified it as Class II and gives the reason as "Integra LifeSciences Corporation received 2 complaints that the USB port on the Licox Pt02 monitor does not consistently provide the user the ability to extract the Pt02 trend data according to the User's Manual for the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0201-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra Licox Pt02 Monitor, Rx Only, Manufacturer: Integra LifeSciences (Ireland) Limited, IDA Business and Technology Park, Sragh, Tullamore, County Offaly, Ireland, Distributed by: Integra LifeSciences Corporation, 311 Enterprise Drive, Plainsboro, NJ 08536
Product code
GWM Device, Monitoring, Intracranial Pressure
Reason for recall
Integra LifeSciences Corporation received 2 complaints that the USB port on the Licox Pt02 monitor does not consistently provide the user the ability to extract the Pt02 trend data according to the User's Manual for the device.
Root cause tag
design
FDA root cause
Device Design
Action
Integra LifeSciences sent an "Urgent Voluntary Medical Device Correction" letter/Return Response Form dated August 18, 2014, to their affected customers via traceable courier service. The letter identified the product the problem and the action needed to be taken by the customer. Complete the attached form. If you do not have an affected monitor, check the box: I do not have any monitors with an affected Serial Number listed below. If you do have affected monitors, check the box: I do have monitors with an affected Serial Number listed below. Circle all the Serial Numbers for the affected monitors you have and record the total quantity. Complete the other information on the form and return the form by email or fax as indicated. Keep a copy of the form for your records. When your form is received, Integra will schedule this service with your facility and provide a temporary loaner monitor to minimize any disruption this action may cause. Should you have any questions, please contact Neuro Customer Service at 1-855-532-1723. We apologize for any inconvenience this may cause and thank you for your cooperation in this effort.
Quantity in commerce
43 (LCX02) and 17 (LCX02R)
Distribution
US Distribution including the states of : CA, DC, FL, IL, IN, MI, MN, NC, NY, PA and TX
Product codes and lots
Catalogue #LCX02 - Serial #'s 2130901032, 2130901050, 2131001062, 2131001063, 2131001086, 2131001087, 2131001102, 2131001110, 2131001111, 2131001115, 2131101124, 2131101125, 2131101127, 2131101129, 2131101134, 2131101138, 2131101151, 2131101154, 2140101181, 2140301222, 2130901017, 2130901019, 2130901022, 2130901025, 2130901034, 2130901035, 2130901051, 2130901055, 2130901056, 2130901058, 2130901059, 2131001072, 2131001076, 2131001077, 2131001081, 2131001088, 2131001093, 2131001108, 2131101126, 2131101129, 2131101122, 2130901028 and 2130901039; Catalogue #LCX02R - Serial #'s 2130901038, 2130901046, 2130901054, 2131201163, 2131201164, 2131201165, 2131201166, 2131201167, 2140101170, 2140101172, 2140101174, 2140101178, 2140101179, 2140201186, 2140201189, 2140201197 and 2140401227.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0201-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0201-2015Class IITerminatedVoluntary: Firm initiated