Z-0085-2015Class I
Baxter 2B8013 Empty INTRA VIA Container with PVC Ports (Sterile, nonpyrogenic fluid path) ,500 ml. Admixture...
Baxter Healthcare Corporation / initiated 2014-09-16 / Terminated / root cause: manufacturing process
Baxter Healthcare Corp began a recall of Baxter 2B8013 Empty INTRA VIA Container with PVC Ports (Sterile, nonpyrogenic fluid path) ,500 ml. Admixture Products and Accessories The lntraVia Empty... on . FDA classified it as Class I and gives the reason as "Particulate matter found inside the fluid path". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0085-2015
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter 2B8013 Empty INTRA VIA Container with PVC Ports (Sterile, nonpyrogenic fluid path) ,500 ml. Admixture Products and Accessories The lntraVia Empty Plastic Container is intended for use in the preparation and administration of drug admixtures.
- Product code
- KPE Container, I.V.
- Reason for recall
- Particulate matter found inside the fluid path.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- Baxter issued a letter to their consignees stating the following: Baxter is requesting that you take the following actions: Action to be taken if product was purchased directly from Baxter. 1. Locate and remove all affected product from all locations in your facility (the product code and lot numbers can be found on the individual product package or shipping carton). 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. The Center for Service can be reached at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time. Please have your Baxter eight-digit ship-to account number ready when calling. 3. Complete the enclosed customer reply form , and return it to Baxter by either tax or scanned e-mail. 4. If you distributed any of the affected products to other facilities, please conduct a recall with your end-user customers in accordance with your customary procedures. Baxter distributed product 2B8011 lot U R 13D 15112 to customers between 4/26/2013 and 6/20/2013. Baxter distributed product 2B8013 lot UR13K1 4095 to customers between 11 /27/2013 and 3/ 10/2014. Action to be taken if product was received from an alternate source. 1. Locate and remove all of the affected products from all locations in your facility (the product code and lot numbers can be found on the individual product package or shipping carton). 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. The Center for Service can be reached at 888-229- 0001 between the hours of 7:00 am and 6:00 pm Central Time. Please have your ship-to account number ready when calling. 3. Follow your supplier's reply and recall process. Please do not return the customer reply form to Baxter.
- Quantity in commerce
- 20448
- Distribution
- Nationwide Distribution including Puerto Rico, and the country of Canada.
- Product codes and lots
- Product code 2B8013 Lot Number: UR13K14095
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0084-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0085-2015 | Class I | Terminated | Voluntary: Firm initiated |