Z-0043-2015Class I

Covidien Medi-Trace Cadence and Kendall Adult Multi-Function Defibrillation Electrodes, Radiotransparent Part...

Medtronic, Inc. / initiated 2014-09-18 / Terminated / root cause: design

Covidien LLC began a recall of Covidien Medi-Trace Cadence and Kendall Adult Multi-Function Defibrillation Electrodes, Radiotransparent Part Number: 22660R on . FDA classified it as Class I and gives the reason as "Covidien Medi-Trace Cadence and Kendall Defibrillation Electrodes Not Compatible for use with Philips model FR3 or FRx AEDs Defibrillators". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0043-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Covidien Medi-Trace Cadence and Kendall Adult Multi-Function Defibrillation Electrodes, Radiotransparent Part Number: 22660R
Product code
DRO Pacemaker, Cardiac, External Transcutaneous (Non-Invasive)
Reason for recall
Covidien Medi-Trace Cadence and Kendall Defibrillation Electrodes Not Compatible for use with Philips model FR3 or FRx AEDs Defibrillators
Root cause tag
design
FDA root cause
Device Design
Action
Covidien notified customers by letter via Federal Express on 9/18/14. The letter requests that they immediately verify that they are not stocking Covidien electrodes for use with Philips FR3 or FRx AED units. A customer reply acknowledgement form is provided to allow Covidien to track the effectiveness of the notification. Covidien is adding a label to all levels of packaging that alerts the user that the product is not compatible with the FR3 and FRx AED units. Distribution partners notified to address product in the distribution chain that is available to be relabeled. For questions customers should call (800) 962-9888, option 8, then extension 2500. For questions regarding this recall call 508-261-8000.
Quantity in commerce
317,550 pairs
Distribution
Worldwide Distribtion-USA (nationwide) including Puerto Rico and the countries of Canada Greece Israel Belgium Finland France Germany Great Britain Ireland Italy Norway Poland Portugal Spain Sweden Switzerland Brazil Chile Philippines Taiwan Australia New Zealand.
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0043-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0043-2015Class ITerminatedVoluntary: Firm initiated
Z-0044-2015Class ITerminatedVoluntary: Firm initiated
Z-0045-2015Class ITerminatedVoluntary: Firm initiated
Z-0046-2015Class ITerminatedVoluntary: Firm initiated
Z-0047-2015Class ITerminatedVoluntary: Firm initiated
Z-0048-2015Class ITerminatedVoluntary: Firm initiated