Z-0046-2015Class I
Covidien Kendall 1710H Multi-Function Defibrillation Electrodes Part Number: 40000006
Medtronic, Inc. / initiated 2014-09-18 / Terminated / root cause: design
Covidien LLC began a recall of Covidien Kendall 1710H Multi-Function Defibrillation Electrodes Part Number: 40000006 on . FDA classified it as Class I and gives the reason as "Covidien Medi-Trace Cadence and Kendall Defibrillation Electrodes Not Compatible for use with Philips model FR3 or FRx AEDs". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0046-2015
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Kendall 1710H Multi-Function Defibrillation Electrodes Part Number: 40000006
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Covidien Medi-Trace Cadence and Kendall Defibrillation Electrodes Not Compatible for use with Philips model FR3 or FRx AEDs
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Covidien notified customers by letter via Federal Express on 9/18/14. The letter requests that they immediately verify that they are not stocking Covidien electrodes for use with Philips FR3 or FRx AED units. A customer reply acknowledgement form is provided to allow Covidien to track the effectiveness of the notification. Covidien is adding a label to all levels of packaging that alerts the user that the product is not compatible with the FR3 and FRx AED units. Distribution partners notified to address product in the distribution chain that is available to be relabeled. For questions customers should call (800) 962-9888, option 8, then extension 2500. For questions regarding this recall call 508-261-8000.
- Quantity in commerce
- 133,930 pairs
- Distribution
- Worldwide Distribtion-USA (nationwide) including Puerto Rico and the countries of Canada Greece Israel Belgium Finland France Germany Great Britain Ireland Italy Norway Poland Portugal Spain Sweden Switzerland Brazil Chile Philippines Taiwan Australia New Zealand.
- Product codes and lots
- Al lot codes
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0043-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0044-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0045-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0046-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0047-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0048-2015 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.