IMRIS Operating Room table (ORT200 and ORT300). Intended for use during diagnostic examinations or surgical...
IMRIS Inc began a recall of IMRIS Operating Room table (ORT200 and ORT300). Intended for use during diagnostic examinations or surgical procedures to support and position a patient on . FDA classified it as Class II and gives the reason as "One of the three pins that is used to hold the ORT200 table alignment block in place on the floor of the MR Suite went missing during testing at a health care facility". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0144-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- IMRIS Imaging, Inc.
- Product
- IMRIS Operating Room table (ORT200 and ORT300). Intended for use during diagnostic examinations or surgical procedures to support and position a patient.
- Product code
- FQO Table, Operating-Room, Ac-Powered
- Reason for recall
- One of the three pins that is used to hold the ORT200 table alignment block in place on the floor of the MR Suite went missing during testing at a health care facility.
- FDA root cause
- Use error
- Action
- Consignees were sent on 10/01/2014 an IMRIS "Urgent Field Safety Notice" letter dated September 25, 2014. The letter described the problem and the product involved in the recall. The letter informed consignees that the pins will be replaced with pins made of non-ferrous material. Customers are to ensure that users are aware of the changes and the existing procedure for table alignment and managing ferrous content in the magnet room. Contact customer support at 1-866-475-0525 for questions.
- Quantity in commerce
- 14
- Distribution
- Worldwide Distribution -- USA, including the states of CA, FL, NH, NC, TN, and UT; and, the countries of AUSTRALIA, CANADA, and JAPAN.
- Product codes and lots
- 10006991, 10006997, 10005201, 10006891, 10004513, 10002692, 10004513, 10002220, 10007376, 10007377, 10002996, and 10002255.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0144-2015 | Class II | Terminated | Voluntary: Firm initiated |