510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0021-2015Class II

Compress Face Reamer

Zimmer GmbH / initiated 2014-09-17 / Terminated / root cause: manufacturing process

Biomet, Inc. began a recall of Compress Face Reamer Product Usage: Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement, tumor resections, revision of... on . FDA classified it as Class II and gives the reason as "Investigation determined that units supplied were missing the 4x21 degree helix angle". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0021-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Compress Face Reamer Product Usage: Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement, tumor resections, revision of previously failed total joint arthroplasty, and trauma.
Product code
HTO Reamer
Reason for recall
Investigation determined that units supplied were missing the 4x21 degree helix angle.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On September 17, 2014, "URGENT MEDICAL DEVICE RECALL NOTICE" was sent to the consignee Biomet Global Supply Chain Center (GSCC) by email and FedEx. GSCC notified all of the customers within Europe, the Middle East, and Asia that received the affected product. The recall notification included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call (574) 372-1570.
Quantity in commerce
4
Distribution
International Distribution in the countries of Netherlands, Europe, the Middle East, and Asia.
Product codes and lots
Catalog numbers: 32-481000  Lot Numbers: 615860, 765230  Catalog Number: 32-481001 Lot Number Identification: 121450, 392410, 765250, 974290

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0021-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0021-2015Class IITerminatedVoluntary: Firm initiated