Z-0214-2015Class III

ADVIA Centaur and ADVIA Centaur XP and ADVIA Centaur CP Cleaning Solution (For in vitro diagnostic use with...

Siemens Healthcare Diagnostics Inc. / initiated 2014-09-12 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc began a recall of ADVIA Centaur and ADVIA Centaur XP and ADVIA Centaur CP Cleaning Solution (For in vitro diagnostic use with the ADVIA Centaur systems) Catalog Number: 112748... on . FDA classified it as Class III and gives the reason as "Stability of the ADVIA Centaur Cleaning Solution is not assured". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0214-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Centaur and ADVIA Centaur XP and ADVIA Centaur CP Cleaning Solution (For in vitro diagnostic use with the ADVIA Centaur systems) Catalog Number: 112748 (Siemens Material Number (SMN): 10310041 The Cleaning Solution is a component of the ADVIA Centaur and ADVIA Centaur XP and ADVIA Centaur CP instruments, and therefore does not have independent Instructions for Use (IFU). Information regarding use of the Cleaning Solution is contained in the owner's manual for the instruments.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Stability of the ADVIA Centaur Cleaning Solution is not assured
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Siemens issued an Urgent Medical Device Recall (UMDR) to customers in the US on September 12, 2014. An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on September 12, 2014. These notices inform the customer of the degradation of hypochlorite in these lots of cleaning solution. Customers were instructed to discontinue use of the affected lots and dispose of the product according to local regulations or procedures. The notices should be reviewed with the facility's laboratory or medical director. Fax-back forms will be used to confirm that the customers received the UMDR. For those customers that do not return a fax-back form, a Siemens representative will follow up with those customers to ensure they received the Urgent Medical Device Recall For questions regarding this recall call 508-668-5000.
Quantity in commerce
4645 (12 pk)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada. Grerat Britain, Saudi Arabia, Nepal, and Viet Nam.
Product codes and lots
Lot Numbers: 459356 Exp. Date: November 2014; 443105 Exp. Date: November 2014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0214-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0214-2015Class IIITerminatedVoluntary: Firm initiated