Z-0120-2015Class II

The Aeroneb Professional Nebulizer System is a portable medical device for multiple patient use that is...

Aerogen Ltd. / initiated 2014-09-02 / Terminated / root cause: manufacturing process

Aerogen Ltd. began a recall of The Aeroneb Professional Nebulizer System is a portable medical device for multiple patient use that is intended to aerosolize physician-prescribed solutions... on . FDA classified it as Class II and gives the reason as "Aerogen has received a small number of complaints in relation to the main power adapter manufactured by FRIWO GmbH and used in the Aeroneb Solo and Pro Nebuliser systems. The complaints received identified that the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0120-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aerogen Ltd.
Product
The Aeroneb Professional Nebulizer System is a portable medical device for multiple patient use that is intended to aerosolize physician-prescribed solutions for inhalation to patients on and off ventilation or other positive pressure breathing assistance. The Aeroneb Professional Nebulizer System is suitable for use in adult and pediatric patients. The Aeroneb Solo Nebulizer System is an iteration of the Aeroneb Professional Nebulizer System which is single patient use. The FRIWO AC/DC adapter is a component of the Aerogen Nebulizer systems which is used to power the Pro and Pro X controllers.
Product code
CAF Nebulizer (Direct Patient Interface)
Reason for recall
Aerogen has received a small number of complaints in relation to the main power adapter manufactured by FRIWO GmbH and used in the Aeroneb Solo and Pro Nebuliser systems. The complaints received identified that the power adapter cover had broken and dislodged from the body of the adapter prior to use.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
URGENT FIELD SAFETY NOTICE letters dated September 2, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to: 1) compare their power adapter date codes against those found in the letter as being subject to this recall; 2) until new power adapters are received, customers may continue to use the recalled power adapters if the customer inspects each power adapter prior to use and finds that there is no evidence of damage or separation of the housing and unplug the adapter after each use; and, 3) contact the Aerogen or appropriate Aerogen sales partner immediately so that the power adapter can be replaced free of charge. The letter further requests that the recall notification be forwarded to any parties that may have received the recalled power adapters. Customers with questions can contact Aerogen at 866-423-76436.
Quantity in commerce
36 power adapters
Distribution
Worldwide Distribution-USA (nationwide) including the states of CA, GA, IL, MS, NC, and TX, and the countries of Australia, Belgium, Canada, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Hungary, India, Ireland, Israel, Italy, Kuwait, Libya, Malaysia, Netherlands, Norway, Poland, Qatar, Russia, South Korea, Singapore, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Product codes and lots
FRIWO AC/DC Adapter Model Number: GPP10;   The FRIWO AC/DC Adapter is supplied with the following Aeroneb Nebulizer systems and their respective model numbers:   1) Model Number: 1079422; Lot Numbers: 9070140620088 , 9070140723088 , 9070140723088;  2) Model Number: 1082693; Lot Numbers: 9073140618134;  3) Model Number: 1082694; Lot Numbers: 9074140530079

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0120-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0118-2015Class IITerminatedVoluntary: Firm initiated
Z-0119-2015Class IITerminatedVoluntary: Firm initiated
Z-0120-2015Class IITerminatedVoluntary: Firm initiated
Z-0121-2015Class IITerminatedVoluntary: Firm initiated