Z-0451-2015Class II

Innova IGS 540 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and...

GE Healthcare / initiated 2014-09-04 / Terminated / root cause: design

GE Healthcare, LLC began a recall of Innova IGS 540 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular... on . FDA classified it as Class II and gives the reason as "Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0451-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Innova IGS 540 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Loss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were sent on 9/4/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref#12218 dated September 4, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. Customers are to perform a system functional check as instructed in the letter. If the check fails, they should contact their local GE Healthcare Service Representative. GE Healthcare will correct all systems at no cost to the customer, and a representative will contact customers to arrange for the correction. For questions they can call US 800-437-1171. For other countries they can call their local GE Healthcare Service Representative.
Quantity in commerce
10 (4 US, 6 OUS).
Distribution
Worldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.
Product codes and lots
Mfg Lot or Serial # System ID 00000642076BU4 303695IR1 00000642193BU7 0004040462 00000646910BU0 314768SM10 00000647792BU1 904UFHNSP 00000646492BU9 XV386041GE8 00000646048BU9 S4160099 00000644611BU6 H2009VAS44 00000646493BU7 H0008VAS06 00000645027BU4 XV106184CX5 00000646491BU1 4184240

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0451-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0451-2015Class IITerminatedVoluntary: Firm initiated
Z-0452-2015Class IITerminatedVoluntary: Firm initiated
Z-0453-2015Class IITerminatedVoluntary: Firm initiated