eFilm Workstation 4.0 and 4.0.1. eFilm Workstation is one of the components of a PACS (Picture Archiving and...
Merge Healthcare, Inc. began a recall of eFilm Workstation 4.0 and 4.0.1. eFilm Workstation is one of the components of a PACS (Picture Archiving and Communications System). Product Usage: eFilm... on . FDA classified it as Class II and gives the reason as "There is an issue related to eFilm Workstation versions 4.0 and 4.0.1 when having multiple studies open and utilizing the thumbnail panel to select multiple series from multiple studies may lead to the selection of an...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0068-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merge Healthcare Incorporated
- Product
- eFilm Workstation 4.0 and 4.0.1. eFilm Workstation is one of the components of a PACS (Picture Archiving and Communications System). Product Usage: eFilm Workstation is a software application that is used for viewing medical images. eFilm Workstation receives digital images and data from various sources (including but not limited to CT,MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). Images are stored, communicated, processed and displayed on the local disc of a workstation and/or across computer networks at distributed locations. Tasks that users may perform when viewing images include, but are not limited to: adjustment of window width and level; image stacking; annotation and measurement of regions of interest; and inversion, rotation, and flips of images. In addition, eFilm Workstation can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- There is an issue related to eFilm Workstation versions 4.0 and 4.0.1 when having multiple studies open and utilizing the thumbnail panel to select multiple series from multiple studies may lead to the selection of an incorrect study.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Merge sent an Important Product Information Notice letter, dated August 22, 2014 to consignees. The letter stated the issue, identified affected product, provided instructions, and told of a product update expected to be available by September 15, 2014. Customers with questions should contact Customer Service at 877-741-5369
- Quantity in commerce
- 2,630
- Distribution
- Worldwide Distribution - US Nationwide and the countries of : Argentina, Australia, Canada, Chile, France, Germany, Hungary, Iceland, Italy, Lithuania, Mexico, New Zealand, Romania, Russia, South Africa, Sweden, and Yemen.
- Product codes and lots
- versions 4.0 and 4.0.1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0068-2015 | Class II | Terminated | Voluntary: Firm initiated |