Z-0948-2015Class II

The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing

Abbott Molecular, Inc. / initiated 2013-11-23 / Terminated / root cause: manufacturing process

Abbott Molecular began a recall of The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing on . FDA classified it as Class II and gives the reason as "The x-drive assembly may fail prematurely. This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0948-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Molecular, Inc.
Product
The Abbott m2000sp is an automated system for performing sample preparation for nucleic acid testing.
Product code
JJH Clinical Sample Concentrator
Reason for recall
The x-drive assembly may fail prematurely. This could cause the manipulator or liquid handling arm to contact an object, shake and drip, or improperly pipet.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Abbott Molecular sent a different Field Correction Recall / Urgent Field Safety Notice letters dated October 24, 2014 to each of three sets of customers. Customers whose instruments' X-Drive Assembly belts were perviously inspected by a Field Service Representative and found to not require replacement were advised to contact their local Abbott Molecular Representative or Abbott Molecular Technical Support at 800-553-7042 x 2 should they hear any unusual noises associated with the LiHa or RoMa movement or misalignment of the LiHa or RoMa. Customers whose instruments' X-Drive Assembly belts were previously inspected and replaced by a Field Service Representative were notifieied that this maintenance activity took place. Customers with questions were instructed to contact Abbott Molelcular. For questions regarding this recall call 224-361-7000. Customers whose instruments' X-Drive Assembly belts have not been inspected were notified that a Field Service Representative is scheduled to visit their facility to inspect and if necessary, replace the X-Drive Assembly belt(s). Customers were advised to contact their local Abbott Molecular Representative should they hear any unusual noises associated with the LiHa or RoMa movement or misalignment of the LiHa or RoMa.
Quantity in commerce
176 devices
Distribution
Worldwide Distribution - USA including CA, FL, GA, IL, MI, NC, NJ, NY, OH, OK, OR, and TX and Internationally to Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, China, Colombia, Costa Rica, Denmark, Dominican Republic, Egypt, Finland, France, Germany, Indonesia, Ireland, Israel, Italy, Kazakhstan, Kenya, Malawi, Mali, Mexico, Morocco, Mozambique, New Zealand, Nigeria, Norway, Pakistan, Poland, Russia, Saudi Arabia, Senegal, South Africa, South Korea, Taiwan, Thailand, Turkey, Ukraine, and United Kingdom.
Product codes and lots
List Number: 09K14-02; Model: E-series; Serial Numbers: 10664 - 10833, 1004, 1005, 10142, 10154, 10171, 10416

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0948-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0948-2015Class IITerminatedVoluntary: Firm initiated