510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0097-2015Class II

Suction canisters labels as follows: 1) DeRoyal(R) SUCTION CANISTER, 1000cc, SEMI-RIGID, REF 71-6510...

DeRoyal Industries, Inc. / initiated 2014-08-22 / Terminated / root cause: manufacturing process

DeRoyal Industries Inc began a recall of Suction canisters labels as follows: 1) DeRoyal(R) SUCTION CANISTER, 1000cc, SEMI-RIGID, REF 71-6510, NON-STERILE, Rx Only. 2) NOVAPLUS(TM) SUCTION CANISTER... on . FDA classified it as Class II and gives the reason as "These lots of suction canisters and lids have an insufficient seal under low or intermittent suction condition". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0097-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DeRoyal Industries, Inc.
Product
Suction canisters labels as follows: 1) DeRoyal(R) SUCTION CANISTER, 1000cc, SEMI-RIGID, REF 71-6510, NON-STERILE, Rx Only. 2) NOVAPLUS(TM) SUCTION CANISTER, 1000 cc, SEMI-RIGID, Cat. No. V71-6510, NON-STERILE, Rx Only general hospital
Product code
KDQ Bottle, Collection, Vacuum
Reason for recall
These lots of suction canisters and lids have an insufficient seal under low or intermittent suction condition.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
DeRoyal sent an Urgent Voluntary Recall letter dated August 22, 2014, to all affected consignees by overnight letter or by email. DeRoyal was to follow up with non-responding accounts before October 1, 2014. Customers were instructed to check their inventory for the affected product. If they have affected product, customers were instructed to destory both the liner and the lid and dispose of it according to their facility's procedures. Customers were instructed to return the included Notice of Destruction to [email protected] even if they no longer have affected product. Customers with questions were instructed to contact their DeRoyal Sales Representative or DeRoyal Customer Service at 1-800-251-9864. For questions regarding this recall call 865-362-2334.
Quantity in commerce
13,900 units
Distribution
Nationwide Distribution including MI, OH, IN, FL, WV, VA, MI, NC, and NY
Product codes and lots
1) DeRoyal(R), REF 71-6510, Lot 36408631  2) NOVAPLUS(TM) Cat. No. V71-6510, Lot 36408622

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0097-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0097-2015Class IITerminatedVoluntary: Firm initiated
Z-0098-2015Class IITerminatedVoluntary: Firm initiated