Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0122-2015Class II

IMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitro...

Siemens Healthcare Diagnostics Inc. / initiated 2014-09-19 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics began a recall of IMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitro diagnostic use with the IMMULITE 1000 Systems Analyzers... on . FDA classified it as Class II and gives the reason as "Customer complaints were received for positive bias reported on the controls used on the IMMULITE /IMMULITE 1000 on the Cortisol Assay reagent lots IMMULITE /IMMULITE 1000 (LKCO1) Lots 381,382 and 383 . Subsequently...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0122-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
IMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitro diagnostic use with the IMMULITE 1000 Systems Analyzers for the quantitative measurement measurement of cortisol (hydrocortisone, Compound F) in serum, as an aid in the clinical assessment of adrenal status.
Product code
CGR Radioimmunoassay, Cortisol
Reason for recall
Customer complaints were received for positive bias reported on the controls used on the IMMULITE /IMMULITE 1000 on the Cortisol Assay reagent lots IMMULITE /IMMULITE 1000 (LKCO1) Lots 381,382 and 383 . Subsequently, the firm confirmed a positive bias on the IMMULITE/IMMULITE 1000 (LKCO1) Lots 381, 382 and 383.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
US Customers were sent an Urgent Medical Device Recall Letter (UMDR2014-09-19, dated 9/19/14) via FedEx on 9/19/14. Outside US Customers were sent an Urgent Field Safety Notice Letter. The letters inform the consignees of the recall of the IMMULITE/IMMULITE 1000 Cortisol (LKCO1) assay, kit lots 381, 382 and 383. Customers were asked to take the following actions: - Please review this letter with your Medical Director. - Discontinue use of and discard the IMMULITE/IMMULITE 1000 Cortisol kit lots listed in the letter. - Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens for reporting to the authorities. --- Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Field Safety Notice letter within thirty (30) days.
Quantity in commerce
US: 772 units, Canada: 43 units
Distribution
Worldwide Distribution: US (nationwide) and internationally to Canada.
Product codes and lots
Lots: 381 (exp. 10/31/2014), 382 (exp. 10/31/2014) and 383 (exp. 11/30/2014)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0122-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0122-2015Class IITerminatedVoluntary: Firm initiated