Z-0146-2015Class II

ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System

Roche Diagnostics GmbH / initiated 2014-09-25 / Terminated / root cause: manufacturing process

Roche Diagnostics Operations, Inc. began a recall of ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System Product Usage: The ACCU-CHEK Combo System is meant only for single-patient... on . FDA classified it as Class II and gives the reason as "The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0146-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
ACCU-CHEK Spirit Combo insulin infusion pump sold as part of Accu-Chek Combo System Product Usage: The ACCU-CHEK Combo System is meant only for single-patient use; the device may not be used for screening or diagnosis of diabetes or for multiple patient use. It is indicated for the treatment of insulin requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger. The ACCU-CHEK Spirit Combo Insulin Pump is intended for the subcutaneous continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin as prescribed by a physician.
Product code
LZG Pump, Infusion, Insulin
Reason for recall
The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the pump prompts customers to confirm the time and date, if the user does not recognize the time and date have returned to default, a shift of their basal rate time block could occur.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Roche sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated September 25, 2014 via UPS Ground (receipt required) to the consignees. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Customer Care Service Center at 1-800-688-4578.
Quantity in commerce
8116
Distribution
US Nationwide Distribution in the states of CA, CT, IA, NV, OH, IN, WI, GA, MO, NY, VA, LA, ME, TX, NC, ND, SD, NJ, CO, AL, SC, IL, KY, FL, OR, MN, MD, MS, AR, WV, WA, TN, CT, ID, KS, MI, PA, UT, MA, RI, NM, OK, DE, AZ, NE, MT, NV, WY, AK, and NH.
Product codes and lots
Model/Catalog/Part Number:  Accu-Chek Combo Kit mg US/English version, 05458544001  Accu-Chek Combo Kit mg US/Spanish version, 055004686001  Accu-Chek Spirit Combo Replacement Pump US/EN, 05391512001  Accu-Chek Spirit Combo Demo Pump- US, 0539175000

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0146-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0146-2015Class IITerminatedVoluntary: Firm initiated