Z-0180-2015Class I

Covidien Puritan Bennett 980 Ventilator, Rx ONLY, Made in Ireland., Covidien LLC, USA. REF 980X1ENDIUU

Medtronic, Inc. / initiated 2014-10-01 / Terminated / root cause: design

Nellcor Puritan Bennett Inc. (dba Covidien LP) began a recall of Covidien Puritan Bennett 980 Ventilator, Rx ONLY, Made in Ireland., Covidien LLC, USA. REF 980X1ENDIUU on . FDA classified it as Class I and gives the reason as "Covidien is recalling certain Puritan Bennett 980 Ventilator Systems due to a blacklight inverter PCBA where a capacitor component can fail, causing a dim but functional screen and thermal odor". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0180-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Covidien Puritan Bennett 980 Ventilator, Rx ONLY, Made in Ireland., Covidien LLC, USA. REF 980X1ENDIUU.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Covidien is recalling certain Puritan Bennett 980 Ventilator Systems due to a blacklight inverter PCBA where a capacitor component can fail, causing a dim but functional screen and thermal odor.
Root cause tag
design
FDA root cause
Device Design
Action
Update: Consignees were notified on 12/11/14 of strategy change after Covidien sent an Urgent Field Safety Corrective Action notification with instructions dated September 2014 via traceable courier to affected consignees. The letter identifed the affected product, problem and actions to be taken. Consignees were instructed to complete the attached verification form and fax it to the Covidien RMS contact indicated on the form. If you need further assistance, contact the Technical Support Department @ 1-800-255-6774, option 4, then option 1.
Quantity in commerce
324 units
Distribution
Worldwide Distribution-US (nationwide) including the states of CA, CO, FL, GA, KY, MA, MN, NC, NY, OH, OK, PA, SC, TN, TX, UT, and WI, and the countries of Canada, Mexico, Saudi Arabia, South Africa and United Arab Emirates.
Product codes and lots
Model number PB980 Ventilator:35B1400143, 35B1400684, 35B1400627, 35B1400708, 35B1400398, 35B1400482,  35B1400491, 35B1400512, 35B1400601, 35B1400121, 35B1400568, 35B1400590, 35B1400595, 35B1400599, 35B1400606, 35B1400106, 35B1400146, 35B1400169, 35B1400226, 35B1400260, 35B1400270, 35B1400263, 35B1400351, 35B1400373, 35B1400400, 35B1400401, 35B1400608, 35B1400089, 35B1400112, 35B1400683, 35B1400685, 35B1400678, 35B1400118, 35B1400114, 35B1400355, 35B1400144, 35B1400300, 35B1400379, 35B1400412, 35B1400472, 35B1400489, 35B1400490, 35B1400492, 35B1400493, 35B1400499, 35B1400500, 35B1400501, 35B1400502, 35B1400507, 35B1400577,  35B1400207, 35B1400211, 35B1400216, 35B1400222, 35B1400225, 35B1400231, 35B1400267, 35B1400312, 35B1400323, 35B1400340, 35B1400383, 35B1400128, 35B1400363, 35B1400459, 35B1400589, 35B1400616,  35B1400637, 35B1400483, 35B1400311, 35B1400447,  35B1400648, 35B1400651, 35B1400654, 35B1400662,  35B1400686, 35B1400692, 35B1400175, 35B1400294, 35B1400436, 35B1400485, 35B1400586, 35B1400644,  35B1400687, 35B1400689, 35B1400694, 35B1400700,  35B1400661, 35B1400676, 35B1400693, 35B1400732,  35B1400734, 35B1400736, 35B1400737, 35B1400742, 35B1400280, 35B1400360, 35B1400471, 35B1400609,  35B1400674, 35B1400688, 35B1400755, 35B1400480,  35B1400675, 35B1400706, 35B1400771, 35B1400667,  35B1400704, 35B1400705, 35B1400727, 35B1400743,  35B1400279, 35B1400285, 35B1400262, 35B1400265, 35B1400259, 35B1400494, 35B1400403, 35B1400779, 35B1400783, 35B1400509, 35B1400724, 35B1400763, 35B1400767, 35B1400566, 35B1400768, 35B1400772,  35B1400791, 35B1400793, 35B1400817, 35B1400525, 35B1400796, 35B1400801, 35B1400809, 35B1400821,  35B1400828, 35B1400333, 35B1400366, 35B1400392,  35B1400440, 35B1400463, 35B1400478, 35B1400308,  35B1400347, 35B1400367, 35B1400396, 35B1400078,  35B1400099, 35B1400130, 35B1400155, 35B1400157, 35B1400158, 35B1400174, 35B1400176, 35B1400178, 35B1400180, 35B1400281, 35B1400315, 35B1400631, 35B1400635, 35B1400625, 35B1400653, 35B1400843, 35B1400847, 35B1400233, 35B1400271, 35B1400337, 35B1400350, 35B1400359, 35B1400368, 35B1400384, 35B1400468, 35B1400486, 35B1400495, 35B1400094, 35B1400236, 35B1400437, 35B1400438, 35B1400479, 35B1400496, 35B1400505, 35B1400506, 35B1400427, 35B1400511, 35B1400325, 35B1400462, 35B1400514, 35B1400536, 35B1400537, 35B1400544, 35B1400066, 35B1400200, 35B1400782, 35B1400829, 35B1400015, 35B1400110, 35B1400516, 35B1400517, 35B1400859, 35B1400013, 35B1400103, 35B1400167, 35B1400124, 35B1400007, 35B1400058, 35B1400083, 35B1400129, 35B1400132, 35B1400006, 35B1400079, 35B1400113, 35B1400116, 35B1400219, 35B1400163, 35B1400543.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0180-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0180-2015Class ITerminatedVoluntary: Firm initiated