Z-0127-2015Class II
Dimension Vista HbA1c
Siemens Healthcare Diagnostics Inc. / initiated 2014-08-14 / Terminated / root cause: design
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista HbA1c on . FDA classified it as Class II and gives the reason as "In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0127-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista HbA1c
- Product code
- LCP Assay, Glycosylated Hemoglobin
- Reason for recall
- In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An Urgent Medical Device Correction, dated August 2014, was provided to all Dimension Vista HbA1c customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers.
- Quantity in commerce
- 20043
- Distribution
- Worldwide Distribution: US (nationwide) and country of: Canada.
- Product codes and lots
- catalog #K3105A, material #10470481, All in-date lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0127-2015 | Class II | Terminated | Voluntary: Firm initiated |