Z-1075-2015Class I
The Plum A+ Infusion System is designed to meet the fluid delivery requirements of today's evolving...
Abbott Medical / initiated 2014-07-02 / Terminated
Hospira Inc. began a recall of The Plum A+ Infusion System is designed to meet the fluid delivery requirements of today's evolving healthcare environments. It is a cassette based... on . FDA classified it as Class I and gives the reason as "One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1075-2015
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- The Plum A+ Infusion System is designed to meet the fluid delivery requirements of today's evolving healthcare environments. It is a cassette based multi-function infusion system. The Plum A+ allows two lines in and one line out. The pump can be used for standard, piggyback, or concurrent delivery. Therapy modes include: Standard Infusions, Multistep Programming, Loading Dose, and Dose Calculation. The Plum A+ is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. Each system includes a pumping module and an assortment of disposable IV sets, optional accessories, and an operator's manual.
- Product code
- FRN Pump, Infusion
- Reason for recall
- One lot of alarm assemblies used in Plum A+ and Plum A+3 infusion pumps may fail to sound at all volume levels.
- FDA root cause
- Component change control
- Action
- Hospira sent an Urgent Medical Device Correction letter dated July 2, 2014, to all affected customers. The letter notified customers that Hospira will be contacting them regarding the completion of an audible alarm test and will replace any alarm assemblies that fail to audible alarm test. Additionally, the letter included instructions for customers to: 1) inform potential users of this notification; 2) complete and return the attached reply form; and, 3) if the products were further distributed, notify those accounts of this recall and ask them to contact Stericycle at 888-912-7350 (Monday - Friday, 8:00 AM - 5:00 PM ET) to receive a reply form. A second URGENT MEDICAL DEVICE CORRECTION letter dated July 16, 2014 was sent to all customers of record because the initial letters may have contained errors in the attached data files listing the potentially affected devices at the customer's' facility. The instructions to the customer were unchanged from the original letter. For questions regarding this recall call 800-441-4100.
- Quantity in commerce
- 190 infusion pumps
- Distribution
- Worldwide Distribution - USA including AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and Brazil, Canada, Colombia, Ecuador, France, Israel, Italy, Mexico, Oman, Peru, Saudi Arabia, Spain, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Product codes and lots
- 1) List Number:12618: Serial Numbers: 14040926 , 14041563 , 14041565 , 14041611 , 14041627 , 14041901 , 14041912 , 14041921 , 14041953 , 14041960 , 14041964 , 14041973 , 14040654 , 14041065 , 14041554 , 14041604 , 14041606 , 14041621 , 14041902 , 14042270 , 14042280 , 14042475 , 14042477 , 14042495 , 14042514 , 14042524 , 14044012 , 14044024 , 14040142 , 14041107 , 14041800 , 14041989 , 14042077 , 14042462 , 14042490 , 14042499 , 14042502 , 14042820 , 14043029 , 14043312 , 14043750 , 14044015 , 14044018 , 14044025 , 14044028 , 14044055 , 14044448 , 14044459 , 14044467 , 14044469 , 14044472 , 14044481 , 14044497 , 14044507 , 14044510 , 14044689 , 14045142 , 14045145 , 14045161 , 14045175 , 14045181 , 14045520 , 14045758 , 14045789 , 14045995 , 14046010 , 14047080 , 14047113 , 14047181 , 14047217 , 14040140 , 14040196 , 14040459 , 14040626 , 14040628 , 14040744 , 14040976 , 14041070 , 14041095 , 14041248 , 14041270 , 14041349 , 14041616 , 14041785 , 14041869 , 14042152 , 14042267 , 14042637 , 14042784 , 14043035 , 14043385 , 14044035 , 14044449 , 14044452 , 14044457 , 14044461 , 14044468 , 14044490 , 14044495 , 14044496 , 14044498 , 14044500 , 14044502 , 14044503 , 14044506 , 14044514 , 14044518 , 14044539 , 14044577 , 14044896 , 14044909 , 14044993 , 14045120 , 14045176 , 14045177 , 14045186 , 14045952 , 14045976 , 14046019 , 14046154 , 14046698 , 14046700 , 14046781 , 14046822 , 14046827 , 14047258 , 14047805 , 18627447 , 18627450 , 18627451 , 18627452 , 18627455 , 18627459 , 18627460 , 18627461 , 18627462 , 18627464 , 18627466 , 18627467 , 18627475 , 18627490 , 18627495 , 18627502 , 18627505 , 18627508 , 18627554 , 18627572 , 18627573 , 18627607 , 18627704 , 18627710 , 18627711 , 18627714 , 18627715 , 18627720 , 18627725 , 18627731 , 18627732 , 18627733 , 18627761 , 18627762 , 18627765 , 18627766 , 18627767 , 18627783 , 18627784 , 18627787 , 18627870 , 18627872 , 18627876 , 18627882 , 18627891 , 18627917 , 18627928 , 18627935 , 18627939 , 18627942 , 18627944 , 18627945 , 18627946 , 18627952 , 18627954 , 18627956 , 18627961 , 18627962 , 14044492 , 18627449 , 18627610 , 18627938 , 18627953
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1071-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1072-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1073-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1074-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-1075-2015 | Class I | Terminated | Voluntary: Firm initiated |