510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0105-2015Class II

Philips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705A...

Philips North America LLC / initiated 2014-10-03 / Terminated / root cause: software

Philips Medical Systems, Inc. began a recall of Philips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705A Intended for: " Monitoring the physiological... on . FDA classified it as Class II and gives the reason as "Philips Intellivue and Avalon Fetal Monitors in time-synchronized automatic/sequence mode, the NBP automatic measurement series is stopped". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0105-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705A Intended for: " Monitoring the physiological parameters of pregnant women " Non-invasive monitoring of fetal heart rates and movements.
Product code
HGM System, Monitoring, Perinatal
Reason for recall
Philips Intellivue and Avalon Fetal Monitors in time-synchronized automatic/sequence mode, the NBP automatic measurement series is stopped
Root cause tag
software
FDA root cause
Software Design Change
Action
Philips Healthcare issued on 10/3/14 the Urgent Medical Device Correction notification/Field Safety Notice. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of a software upgrade free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for service. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: During the interim period until the SW is upgraded please make sure, if the NBP repetition interval needs to be increased (e.g. from 5 to 15 minutes), to restart the automatic NBP measurement. Please review this information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication.
Quantity in commerce
118 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA AUSTRIA BAHRAIN BELGIUM CHINA COLOMBIA CZECH REPUBLIC DENMARK FINLAND FRANCE GERMANY HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN KAZAKHSTAN NETHERLANDS NEW ZEALAND NICARAGUA NIGERIA NORWAY OMAN POLAND ROMANIA RUSSIA SINGAPORE SLOVENIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UKRAINE and UNITED KINGDOM.
Product codes and lots
software revision J.30.58:

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0105-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0104-2015Class IITerminatedVoluntary: Firm initiated
Z-0105-2015Class IITerminatedVoluntary: Firm initiated