Philips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705A...
Philips Medical Systems, Inc. began a recall of Philips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705A Intended for: " Monitoring the physiological... on . FDA classified it as Class II and gives the reason as "Philips Intellivue and Avalon Fetal Monitors in time-synchronized automatic/sequence mode, the NBP automatic measurement series is stopped". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0105-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Avalon Monitors with software revision J.30.58: Model Product FM20 M2702A ; FM30 M2703A ; FM50 M2705A Intended for: " Monitoring the physiological parameters of pregnant women " Non-invasive monitoring of fetal heart rates and movements.
- Product code
- HGM System, Monitoring, Perinatal
- Reason for recall
- Philips Intellivue and Avalon Fetal Monitors in time-synchronized automatic/sequence mode, the NBP automatic measurement series is stopped
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- Philips Healthcare issued on 10/3/14 the Urgent Medical Device Correction notification/Field Safety Notice. The letter informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of a software upgrade free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for service. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: During the interim period until the SW is upgraded please make sure, if the NBP repetition interval needs to be increased (e.g. from 5 to 15 minutes), to restart the automatic NBP measurement. Please review this information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication.
- Quantity in commerce
- 118 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of AUSTRALIA AUSTRIA BAHRAIN BELGIUM CHINA COLOMBIA CZECH REPUBLIC DENMARK FINLAND FRANCE GERMANY HONG KONG INDIA IRELAND ISRAEL ITALY JAPAN KAZAKHSTAN NETHERLANDS NEW ZEALAND NICARAGUA NIGERIA NORWAY OMAN POLAND ROMANIA RUSSIA SINGAPORE SLOVENIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UKRAINE and UNITED KINGDOM.
- Product codes and lots
- software revision J.30.58:
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0104-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0105-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
