Z-0424-2015Class II

Trabecular Metal Tibial Impactor Replacement Pads

Zimmer GmbH / initiated 2014-10-14 / Terminated / root cause: labeling and UDI

Zimmer, Inc. began a recall of Trabecular Metal Tibial Impactor Replacement Pads. Product Usage: TM Tibial Impactor pads are used in conjunction with the TM Tibial Impactor to impact the... on . FDA classified it as Class II and gives the reason as "To provide clarifying instructions relating to the assembly/disassembly of the Trabecular Metal Tibial Impactor with/from TM tibial Impactor Pads due to complaints regarding Tibal Impactor pad fracture". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0424-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Trabecular Metal Tibial Impactor Replacement Pads. Product Usage: TM Tibial Impactor pads are used in conjunction with the TM Tibial Impactor to impact the tibial tray onto the tibia.
Product code
HWA Impactor
Reason for recall
To provide clarifying instructions relating to the assembly/disassembly of the Trabecular Metal Tibial Impactor with/from TM tibial Impactor Pads due to complaints regarding Tibal Impactor pad fracture.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On 10/14/2014, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected distributors and hospital Risk Managers. All distributors were notified via electronic mail, and all hospital Risk Managers and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-877-946-2761.
Quantity in commerce
4961 units
Distribution
Worldwide Distribution - US Nationwide in the state of AL, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, ND, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, WI, NC, TN, OK, AZ, UT, AK, PR, DC. and the countries of: Germany, China, Canada, Australia, Japan, United Arab Emirates, Austria, Belgium, Switzerland, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Iceland, Italy, Malta, Netherlands, Norway, Poland, Russian Fed., Saudi Arabia, Sweden, Slovakia, and South Africa.
Product codes and lots
Item 00-5953-056-06; all lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0424-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0423-2015Class IITerminatedVoluntary: Firm initiated
Z-0424-2015Class IITerminatedVoluntary: Firm initiated