Z-0086-2015Class II

Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography...

Siemens Medical Solutions USA, Inc. / initiated 2014-09-18 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic... on . FDA classified it as Class II and gives the reason as "The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0086-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.
Product code
MUE Full Field Digital, System, X-Ray, Mammographic
Reason for recall
The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i
Root cause tag
manufacturing process
FDA root cause
Process control
Action
A safety advisory notice, dated September 18, 2014, was sent to end users that identified the product, problem, and actions to be taken.
Quantity in commerce
3
Distribution
US Distribution in states of: NY, VA, and MA.
Product codes and lots
material #10140000, with serial numbers: 5018 3849 5042

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0086-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0086-2015Class IITerminatedVoluntary: Firm initiated