Z-0003-2015Class II

Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic...

Siemens Medical Solutions USA, Inc. / initiated 2014-09-02 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic... on . FDA classified it as Class II and gives the reason as "Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0003-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens AXIOM Luminos dRF Max systems The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a Safety Advisory Notice, dated September 2, 2014, to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disable the automatic RIS worklist update and perform this procedure manually. Customers should disable the Measurement Field Display and Collimator Display by setting the display time for both setting to 0 sec (if editable). Customers should forward the safety notice to the new owner and inform Siemens of the new owner where possible. Customers with questions should call 610-219-2119 or 610-219-4834. For questions regarding this recall call 610-219-6300.
Quantity in commerce
1
Distribution
Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.
Product codes and lots
Siemens AXIOM Luminos dRF Max system material #s: 10762471 with serial number: 5037.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0003-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0002-2015Class IITerminatedVoluntary: Firm initiated
Z-0003-2015Class IITerminatedVoluntary: Firm initiated
Z-0004-2015Class IITerminatedVoluntary: Firm initiated
Z-0005-2015Class IITerminatedVoluntary: Firm initiated