Z-0004-2015Class II

Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system...

Siemens Medical Solutions USA, Inc. / initiated 2014-09-02 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray... on . FDA classified it as Class II and gives the reason as "Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0004-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Uroskop Omnia Max system The Uroskop Omnia is a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: "Querying and retrieving patient history information and/or previous diagnosis and images from other modalities; "X-ray examinations of the urogenital area "Ultrasound examinations "Endourological interventions "Percutaneous interventions "Laparoscopy "Application of fistula "Simple procedures "Extracorporeal shock wave lithotripsy "Uroflow/urodynamics "Pediatric radiological and therapeutic applications
Product code
KPR System, X-Ray, Stationary
Reason for recall
Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a Safety Advisory Notice, dated September 2, 2014, to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disable the automatic RIS worklist update and perform this procedure manually. Customers should disable the Measurement Field Display and Collimator Display by setting the display time for both setting to 0 sec (if editable). Customers should forward the safety notice to the new owner and inform Siemens of the new owner where possible. Customers with questions should call 610-219-2119 or 610-219-4834. For questions regarding this recall call 610-219-6300.
Quantity in commerce
2
Distribution
Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.
Product codes and lots
Siemens Uroskop Omnia Max system, material #s: 10762473, serial numbers: 4014 4015

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0004-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0002-2015Class IITerminatedVoluntary: Firm initiated
Z-0003-2015Class IITerminatedVoluntary: Firm initiated
Z-0004-2015Class IITerminatedVoluntary: Firm initiated
Z-0005-2015Class IITerminatedVoluntary: Firm initiated