Z-0005-2015Class II

Siemens Luminos Agile Max system The Luminos Agile is intended to be used as a universal imaging system for...

Siemens Medical Solutions USA, Inc. / initiated 2014-09-02 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens Luminos Agile Max system The Luminos Agile is intended to be used as a universal imaging system for radiographic and fluoroscopic studies. Using either... on . FDA classified it as Class II and gives the reason as "Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0005-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Luminos Agile Max system The Luminos Agile is intended to be used as a universal imaging system for radiographic and fluoroscopic studies. Using either film cassettes or a digital mobile flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. Luminos Agile is applicable to emergency treatment on an outpatient basis, as well as for bedside examinations.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Two malfunctions posing potential risk to patients were identified with Siemens Ysio Max, AXIOM Luminos dRF Max, Uroskop Omnia Max, and Luminos Agile Max systems with a specific set of serial numbers that will cause lost images due to an automatic Radiology Information System (RIS) worklist update, thus causing examinations to be repeated. Image label may also be displayed incorrectly after images
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a Safety Advisory Notice, dated September 2, 2014, to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disable the automatic RIS worklist update and perform this procedure manually. Customers should disable the Measurement Field Display and Collimator Display by setting the display time for both setting to 0 sec (if editable). Customers should forward the safety notice to the new owner and inform Siemens of the new owner where possible. Customers with questions should call 610-219-2119 or 610-219-4834. For questions regarding this recall call 610-219-6300.
Quantity in commerce
5
Distribution
Nationwide Distribution - US including UT, MO, CO, MA, MN, MS, OH, SC, MI, and NY.
Product codes and lots
Siemens Luminos Agile Max system material #s: 10762472, with serial numbers: 61013 61015 61011 61014 61017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0005-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0002-2015Class IITerminatedVoluntary: Firm initiated
Z-0003-2015Class IITerminatedVoluntary: Firm initiated
Z-0004-2015Class IITerminatedVoluntary: Firm initiated
Z-0005-2015Class IITerminatedVoluntary: Firm initiated