Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0133-2015Class II

SAM Junctional Tourniquet (SJT) Auxiliary, Model SJT 102 and SJT 112. The SJT models specific intended uses...

Sam Medical Products / initiated 2014-08-14 / Terminated / root cause: design

Sam Medical Products began a recall of SAM Junctional Tourniquet (SJT) Auxiliary, Model SJT 102 and SJT 112. The SJT models specific intended uses are to control difficult bleeds in the inguinal... on . FDA classified it as Class II and gives the reason as "SAM Junctional Tourniquet (SJT) Auxiliary strap, model SJT 102 and SJT 112, is recalled because the snap hook on auxiliary strap broke during training demonstration". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0133-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Sam Medical Products
Product
SAM Junctional Tourniquet (SJT) Auxiliary, Model SJT 102 and SJT 112. The SJT models specific intended uses are to control difficult bleeds in the inguinal area and to immobilize a pelvic fracture.
Product code
DXC Clamp, Vascular
Reason for recall
SAM Junctional Tourniquet (SJT) Auxiliary strap, model SJT 102 and SJT 112, is recalled because the snap hook on auxiliary strap broke during training demonstration.
Root cause tag
design
FDA root cause
Component design/selection
Action
SAM Medical Products issued the press release on August 29, 2014 and the link is http://www.ireachcontent.com/news-releases/sam-medical-products-implements-voluntary-recall-of-its-accessory-axilla-strap-for-the-sam-junctional-tourniquet-273239381.html. SAM Medical Products sent the "URGENT: MEDICAL DEVICE RECALL" letter, dated September 10, 2014, to their consignees. The 5 additional lot numbers (7/1/14, 7/2/14, 8/4/14, 8/7/14, and 8/11/14) that were erroneously placed on the recall notification letter and they are NOT part of the recall. SAM Medical Distributors will need to notify their customers via this letter and the included FAX Back document of the actions being taken as well as ship the replacement straps to individual customers. Distributors will document and report to SAM Medical that all customers have been notified and replacement straps delivered to these customers. Distributors will notify SAM Medical that the replaced straps have been destroyed. This information will be documented on the included FAX Back document and then returned to SAM Medical (via FAX, e-mail, or mail, per instructions on the FAX Back document). Customers will receive one complete Auxiliary Accessory Strap at no charge with a reinforced metal clip as a replacement. Once received, the existing strap should be removed from the SJT and replaced with the new strap. Customers/End Users need to complete and return the FAX Back document to the distributor from whom the FAX Back was sent. Customer can contact SAM Medical Customer Service at: -Email [email protected] -Phone (503) 639-5474, Monday through Friday, 7:30-4:30, Pacific Time -Toll free l-800-818-4726 -Website: vww.sammedical.com
Quantity in commerce
total 3633 units (3342 units in US and 291 units Outside US)
Distribution
Worldwide Distribution: US (nationwide) in states of Ohio, Oregon, New Mexico, North Carolina, Texas and to the following countries: Australia, Canada, Hong Kong, Iceland, Japan, Lebanon, Singapore, Thailand, and United Kingdom.
Product codes and lots
SJT 112 Lot number (represented as date codes): 6/5/14;  SJT 102 Lot numbers (represented as date codes):  3/29/2013, 5/15/2013, 6/11/2013, 7/16/2013, 8/27/2013, 9/3/2013, 9/10/2013, 11/4/2013, 12/9/2013, 12/11/2013, 1/10/2014, 1/30/2014, 2/25/2014, 3/4/2014, 3/5/2014, 3/19/2014, 3/21/2014, 3/24/2014, 3/31/2014, 4/1/2014, 4/2/2014, 4/4/2014, 4/7/2014, 4/9/2014, 4/11/2014, 4/22/2014, 4/28/2014, 4/29/2014, 5/5/2014, and 6/12/2014.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0133-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0133-2015Class IITerminatedVoluntary: Firm initiated