510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0124-2015Class II

Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer...

Cook Medical LLC / initiated 2014-09-26 / Terminated / root cause: packaging

Cook Medical Incorporated began a recall of Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer needle that consists of a cutting cannula and a trocar... on . FDA classified it as Class II and gives the reason as "MEDINVENTS, the manufacturer of the Spirotome" Soft-Tissue Biopsy Needle Set distributed by Cook Medical, has initiated a voluntary recall of these products. The recall has been initiated due to customer complaints that...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0124-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Medical LLC
Product
Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer needle that consists of a cutting cannula and a trocar stylet, a helical-tip biopsy needle that has been sized to fit through the lumen of the cutting cannula. The Spirotome Biopsy Needle Set is intended for percutaneous acquisition of soft tissue specimens for laboratory analysis.
Product code
KNW Instrument, Biopsy
Reason for recall
MEDINVENTS, the manufacturer of the Spirotome" Soft-Tissue Biopsy Needle Set distributed by Cook Medical, has initiated a voluntary recall of these products. The recall has been initiated due to customer complaints that the protective caps were dislodged in the pouch. Further investigation disclosed potential dislodging of protective caps on the needle tips on other distributed products.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
On 9/26/2014, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected distributors and institutions Risk Managers/Recall Administration with instructions for returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call Customer Relations at 1-800-457-4500 or 1-812-339-2235.
Quantity in commerce
1239 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of Arizona, California, Colorado, Hawaii, Iowa, Illinois, Massachusetts, Maryland, Michigan, Missouri, New Mexico, Pennsylvania, Texas, Virginia, Vermont, and Wisconsin, and the countries of AT; BE; CH; CY; CZ; DE; DK; ES; FI; FR; GB; HU; IC; IE; IS; IT; LB; LT; LU; NL; NO; PT; QA; SE; TR; and ZA.
Product codes and lots
Model Numbers: SS-08-06; SS-08-10; SS-08-15; SS-10-06; SS-10-10;  SS-10-15; SS-14-06; SS-14-10; SS-14-15.   Lot numbers: P030214C; P020413A; P190813B; P020413B; P030214A; P020413F; P300913A; P020413D; P190813D; P030214B; P020413E; P190813F; P300913B; P020413C; P190813E.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0124-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0124-2015Class IITerminatedVoluntary: Firm initiated