Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer...
Cook Medical Incorporated began a recall of Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer needle that consists of a cutting cannula and a trocar... on . FDA classified it as Class II and gives the reason as "MEDINVENTS, the manufacturer of the Spirotome" Soft-Tissue Biopsy Needle Set distributed by Cook Medical, has initiated a voluntary recall of these products. The recall has been initiated due to customer complaints that...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0124-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Medical LLC
- Product
- Spirotome" Soft-Tissue Biopsy Needle Set. The Spirotome Biopsy Needle Set includes a two-part introducer needle that consists of a cutting cannula and a trocar stylet, a helical-tip biopsy needle that has been sized to fit through the lumen of the cutting cannula. The Spirotome Biopsy Needle Set is intended for percutaneous acquisition of soft tissue specimens for laboratory analysis.
- Product code
- KNW Instrument, Biopsy
- Reason for recall
- MEDINVENTS, the manufacturer of the Spirotome" Soft-Tissue Biopsy Needle Set distributed by Cook Medical, has initiated a voluntary recall of these products. The recall has been initiated due to customer complaints that the protective caps were dislodged in the pouch. Further investigation disclosed potential dislodging of protective caps on the needle tips on other distributed products.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- On 9/26/2014, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected distributors and institutions Risk Managers/Recall Administration with instructions for returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call Customer Relations at 1-800-457-4500 or 1-812-339-2235.
- Quantity in commerce
- 1239 units
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of Arizona, California, Colorado, Hawaii, Iowa, Illinois, Massachusetts, Maryland, Michigan, Missouri, New Mexico, Pennsylvania, Texas, Virginia, Vermont, and Wisconsin, and the countries of AT; BE; CH; CY; CZ; DE; DK; ES; FI; FR; GB; HU; IC; IE; IS; IT; LB; LT; LU; NL; NO; PT; QA; SE; TR; and ZA.
- Product codes and lots
- Model Numbers: SS-08-06; SS-08-10; SS-08-15; SS-10-06; SS-10-10; SS-10-15; SS-14-06; SS-14-10; SS-14-15. Lot numbers: P030214C; P020413A; P190813B; P020413B; P030214A; P020413F; P300913A; P020413D; P190813D; P030214B; P020413E; P190813F; P300913B; P020413C; P190813E.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0124-2015 | Class II | Terminated | Voluntary: Firm initiated |