Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part...
Invacare Corporation began a recall of Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of... on . FDA classified it as Class II and gives the reason as "It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0577-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Invacare GmbH
- Product
- Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.
- Product code
- CAW Generator, Oxygen, Portable
- Reason for recall
- It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Invacare sent an Urgent - Medical Device Recall letter to customers, dated October 2014. The letter identified the affected product as well as the problem involved. The letter instructed customers to examine their device and confirm that it has an affected serial number, or if service was conducted on the device using IRC5O2V Service Part Number 1165099 with the affected date codes. Customers were asked to immediately contact their dealer or service provider at the number listed, so that a replacement can provided. A service provider or dealer should be contacted with questions regarding the instructions.
- Quantity in commerce
- 5,963
- Distribution
- Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
- Product codes and lots
- The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0577-2015 | Class II | Terminated | Voluntary: Firm initiated |