Z-0162-2015Class II
Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package...
Merit Medical Systems, Inc. / initiated 2014-09-26 / Terminated / root cause: labeling and UDI
Merit Medical Systems, Inc. began a recall of Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package is unopened or undamaged on . FDA classified it as Class II and gives the reason as "These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0162-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package is unopened or undamaged.
- Product code
- KRA Catheter, Continuous Flush
- Reason for recall
- These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F.
- Root cause tag
- labeling and UDI
- FDA root cause
- Mixed-up of materials/components
- Action
- Consignees were notified via letter on September 26, 2014. Written notification included product identification information, instruction to immediately quarantine any devices and discontinue use, ensure all personnel to whom devices were distributed are made aware of this field action, and instructions to contact their Merit representative. An email notification was sent to US Merit sales representatives with accounts that had received affected product. The affected sales representatives were provided a Customer Response Form applicable to their account, a representative copy of communications sent directly to their affected accounts, and a copy of the Frequently Asked Questions. Merit sales representatives were instructed to visit the account and assist with the return of the affected units back to the Merit Field Assurance Department.
- Quantity in commerce
- 56 units
- Distribution
- Nationwide and to India. No govt/military/VA consignees.
- Product codes and lots
- Catalog No: 28MC24130ST, Lot H665694, H669824, Exp 2017 and Catalog No: 28MC24150SN, Lot H665696, Exp 7/31/2014
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0162-2015 | Class II | Terminated | Voluntary: Firm initiated |