510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0162-2015Class II

Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package...

Merit Medical Systems, Inc. / initiated 2014-09-26 / Terminated / root cause: labeling and UDI

Merit Medical Systems, Inc. began a recall of Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package is unopened or undamaged on . FDA classified it as Class II and gives the reason as "These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0162-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package is unopened or undamaged.
Product code
KRA Catheter, Continuous Flush
Reason for recall
These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
Consignees were notified via letter on September 26, 2014. Written notification included product identification information, instruction to immediately quarantine any devices and discontinue use, ensure all personnel to whom devices were distributed are made aware of this field action, and instructions to contact their Merit representative. An email notification was sent to US Merit sales representatives with accounts that had received affected product. The affected sales representatives were provided a Customer Response Form applicable to their account, a representative copy of communications sent directly to their affected accounts, and a copy of the Frequently Asked Questions. Merit sales representatives were instructed to visit the account and assist with the return of the affected units back to the Merit Field Assurance Department.
Quantity in commerce
56 units
Distribution
Nationwide and to India. No govt/military/VA consignees.
Product codes and lots
Catalog No: 28MC24130ST, Lot H665694, H669824, Exp 2017 and Catalog No: 28MC24150SN, Lot H665696, Exp 7/31/2014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0162-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0162-2015Class IITerminatedVoluntary: Firm initiated