Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0163-2015Class II

Bard Dimension Stone Basket

C.R. Bard Inc / initiated 2014-10-07 / Terminated / root cause: sterility

C.R. Bard, Inc. began a recall of Bard Dimension Stone Basket Product Usage: The Bard Dimension Stone basket is a teardrop-shaped basket with the ability to capture stones by, opening and... on . FDA classified it as Class II and gives the reason as "The product is non-sterile". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0163-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.R. Bard Inc
Product
Bard Dimension Stone Basket Product Usage: The Bard Dimension Stone basket is a teardrop-shaped basket with the ability to capture stones by, opening and closing the basket and articulating or moving the basket from side to side. The device consists of 3 main parts; handle, shaft and basket. It is intended to use in the endoscopic removal of renal and ureteral stones.
Product code
FFL Dislodger, Stone, Basket, Ureteral, Metal
Reason for recall
The product is non-sterile.
Root cause tag
sterility
FDA root cause
Release of Material/Component prior to receiving test results
Action
Bard sent an Urgent: Medical Device Recall letter to each customer on 10/7/2014 with proof of delivery notice. Wholesalers were asked to conduct a sub-recall. Letter identify the affected product, problem and actions to be taken. Customers were instructed to examine their inventory and quarantine product subject to recall, do not use or further distribute any of the affected product, remove and return affected product to Bard Medical Division, and if the product have been further distributed identify your customers and notify them at once of the product recall. Customers were instructed to complete the enclosed Recall & Effectiveness Check Form and fax to 1-770-784-6469. For questions call 1-770-784-6471.
Quantity in commerce
86 units
Distribution
Worldwide Distribution in the states of AL, FL, NJ, NY, NC, PA, SC, TX, WI, and countries of Germany, Italy, Kuwait, and Luxembourg.
Product codes and lots
Product code: 042316, Lot number: BMXJM031

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0163-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0163-2015Class IITerminatedVoluntary: Firm initiated