Z-0196-2015Class II

The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower...

STERIS Corporation / initiated 2014-10-10 / Terminated / root cause: design

Steris Corporation began a recall of The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation on . FDA classified it as Class II and gives the reason as "The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0196-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.
Product code
GDC Table, Operating-Room, Electrical
Reason for recall
The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.
Root cause tag
design
FDA root cause
Device Design
Action
STERIS will mail Consignee Notification Letters to the affected Consignees via FedEx with tracking numbers for delivery confirmation. The Consignee Notification mailing will initiate the week of October 13, 2014.
Quantity in commerce
999
Distribution
US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.
Product codes and lots
5085 and 5085SRT

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0196-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0196-2015Class IITerminatedVoluntary: Firm initiated