The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower...
Steris Corporation began a recall of The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation on . FDA classified it as Class II and gives the reason as "The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0196-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation.
- Product code
- GDC Table, Operating-Room, Electrical
- Reason for recall
- The 5085 and 5085SRT stainless steel column shrouds may become damaged as a result of misalignment during extreme table articulation, excessive side force on the table column, or pressure on the base of the table. Severe damage to the shrouds may prevent the user from raising or lowering the surgical table.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- STERIS will mail Consignee Notification Letters to the affected Consignees via FedEx with tracking numbers for delivery confirmation. The Consignee Notification mailing will initiate the week of October 13, 2014.
- Quantity in commerce
- 999
- Distribution
- US distribution: AL, AZ, AR, CA, CO, CN, DC, FL, GA, HI, ID, IL, IN, IA, KN, KY, LA, ME, MD, MI, MN, MS, MO, NB, NV, NY, NC, ND, OH, OK, OR, PN, RI, SC, SD, TN, TX, VA, WV, WI. Foriegn: Austrialia, Brazil, Canada, China, Ecuador, India, Korea, Mexico, Pnama, and Thailand.
- Product codes and lots
- 5085 and 5085SRT
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0196-2015 | Class II | Terminated | Voluntary: Firm initiated |