Z-0854-2015Class II

Artis zee/ zeego systems, model numbers: 10094135, 10094137, 10094139, 10094141, 10280959: dedicated...

Siemens Medical Solutions USA, Inc. / initiated 2014-09-17 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Artis zee/ zeego systems, model numbers: 10094135, 10094137, 10094139, 10094141, 10280959: dedicated angiography systems developed for single and biplane... on . FDA classified it as Class II and gives the reason as "The Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0854-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis zee/ zeego systems, model numbers: 10094135, 10094137, 10094139, 10094141, 10280959: dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
The Small Focus of a 3 focal spot x-ray tube may fail and further release of radiation will not be possible. The Large Focus and Micro Focus focal spots will still be operational without any restriction. Manual switchover is needed, resulting in a short interruption of the procedure.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Planned action (s) to repair defect or to bring product into compliance for Siemens Medical Solutions Inc. concerning the corrective action plan (CAP) for the Model(s) Artis Zee/Zeego Stationary Angiographic X-ray Systems. 1.Notifying the customer of the defect. 2.Implementing a corrective software patch to protect the system from damage in the event of hardware failure 3. Scheduling a service visit from a Siemens representative to perform a hardware replacement. 4. Replacing the non-conforming X-ray tube with one manufactured after the nonconformity issue was resolved. For further questions please call (610) 219-6300.
Quantity in commerce
44
Distribution
USA Distribution to the states of MO, MI, NY, OK, CA, LA, CT, IN, VA, MD, WA, HI, ME, TX, NY, OH, PA and CO.
Product codes and lots
model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, with serial numbers: 160348 146643 147129 146630 147208 146972 153707 147378 146742 157650 146986 147098 157696 153946 146940 147144 147029 160384 153884 147009 153628 153706 146729 147155 153790 146896 153757 136503 147372 136345 153762 146883 160421 153195 146427 153834 135927 154157 135862 153998 153257 153220 153682 160341

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0854-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0854-2015Class IITerminatedVoluntary: Firm initiated