Artis zeego systems, model number 10280959, with serial numbers 1000-1599 : Artis zeego is a dedicated...

Siemens Medical Solutions USA, Inc. / initiated 2014-09-23 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Artis zeego systems, model number 10280959, with serial numbers 1000-1599 : Artis zeego is a dedicated angiography system developed for single and biplane... on . FDA classified it as Class II and gives the reason as "A cable inside the C-arm systems may have been routed in an inappropriate manner. Improper routing of the cable may result in increased wear over time. Without additional measures, the potential exists for a cable...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0183-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis zeego systems, model number 10280959, with serial numbers 1000-1599 : Artis zeego is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
A cable inside the C-arm systems may have been routed in an inappropriate manner. Improper routing of the cable may result in increased wear over time. Without additional measures, the potential exists for a cable inside the C-arm to break which could result in restricted functionality or system failure and an ongoing procedure could be terminated. Additionally, a subgroup of affected C-arm systems will receive a replacement of the cable harness.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent an " Important Safety Information " letter dated September 23, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Siemens will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. Our service organization will contact with you to schedule an appointment to perform the corrective action. Please feel free to contact our service organization for an earlier appointment. Please observe this safety notice and comply with the corresponding measure until the update has been fully completed. For further questions, please call (610) 219-6300.
Quantity in commerce
164
Distribution
Nationwide Distribution including Puerto Rico.
Product codes and lots
model number 10280959, with serial numbers 1000-1599

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0183-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0183-2015Class IITerminatedVoluntary: Firm initiated