510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0205-2015Class II

Medtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K

Medtronic, Inc. / initiated 2014-10-13 / Terminated / root cause: manufacturing process

Medtronic Vascular began a recall of Medtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K on . FDA classified it as Class II and gives the reason as "Two reports have been received on the 6F launcher Guide Catheter lot (Model Number LA6JR40K; lot No 0007244836} for incorrect usable length of the Guide Catheter. It was reported that the catheters had a usable length...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0205-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K
Product code
DQY Catheter, Percutaneous
Reason for recall
Two reports have been received on the 6F launcher Guide Catheter lot (Model Number LA6JR40K; lot No 0007244836} for incorrect usable length of the Guide Catheter. It was reported that the catheters had a usable length of 90 cm instead of 55 cm, which was specified on the packaging label. Usable length is defined as the total length of catheter shaft from the distal end of the strain relief to the
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Beginning October 13, 2014, Medtronic field representatives will being hand delivery of the customer communication to affected customers, notifying them of the issue, and requesting return of unaffected product. Customers will be asked to sign a certificate confirming they have received the notification and have taken the requested action. 1. Identify and quarantine all listed, unused product in your inventory. 2. Return all listed product in your inventory to Medtronic. Contact Customer Service at 888-283-7868 to initiate a product return and credit. Your Medtronic sales representative can assist you in the return of this product as necessary. 3. Complete the attached Customer Confirmation Certificate and email it to [email protected] or fax it to Medtronic at 651-367-0612 to the attention of Customer Focused Quality.
Quantity in commerce
11 devices
Distribution
US Distribution including the states of NC, IA, IL and KS.
Product codes and lots
Lot Number: 0007244836 MODEL: LA6JR40K

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0205-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0205-2015Class IITerminatedVoluntary: Firm initiated