Z-0529-2015Class II

Soft Contact Lens

COOPERVISION Manufacturing Puerto Rico, LLC / initiated 2014-10-28 / Terminated / root cause: manufacturing process

Cooper Vision Caribbean Corp. began a recall of Soft Contact Lens on . FDA classified it as Class II and gives the reason as "Portions of the lots may contain units with an incorrect lens axis condition". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0529-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
COOPERVISION Manufacturing Puerto Rico, LLC
Product
Soft Contact Lens
Product code
LPM Lenses, Soft Contact, Extended Wear
Reason for recall
Portions of the lots may contain units with an incorrect lens axis condition.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
This recall is being initiated to the wholesale/distributor and Eye Care Practitioner level. Written communications will be forwarded to all first level consignees by overnight mailin..9. Receipt" tracking and follow up with consignees will be conducted according to the Effectiveness Check noted below. . - . The 0ff Axis condition is ~onsistent with a situation in which use of, or exposure to, the product is unlikely to cause adverse health consequences. A user may notice blurred vision due to the axis being out of specification but removal of the lens effectively reduces/eliminates any impairment. There is currently no indication that use of a lens with an Off Axis condition will create an injury requiring medical intervention. No permanent impairment is anticipated with or without medical intervention. Based on the analysis " described in the attached Health Hazard Evaluation, CooperVision will initiate a Class Ill recall. As noted by regulation, this classification is consistent with a situation in which use of, or exposure to, the product is unlikely .to cause adverse health consequences. Our evaluation of the Off Axis condition indicates it is unlikely to cause an adverse health reaction. Based on a Class Ill determination, the recall shall be conducted to the Wholesale level with a planned Effectiveness Check at Level B - 50% of Consignees to be contacted. Public Warning- in light of the low risk and proposed recall classification; a public warning is not warranted at this time. Distribution includes: " North America: Canada, USA " Latin America: Argentina, Brasil, Chile, Colombia, Ecuador, Guatemala, Uruguay, Venezuela, Mexico " Europe: Belgium, Denmark, Germany, France, Netherlands, Poland, Spain, Sweden, Switzerland " Australia Returned lenses will be quarantined from distribution and destroyed as required. Actions and Dates: " Notify FDA through 806 Form on 10/23/14 " Initiate External communications starting 10/28/2014 " C
Quantity in commerce
1170
Distribution
US Nationwide Distribution.
Product codes and lots
Lots # 8246525202 & 8255525096

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0529-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0529-2015Class IITerminatedVoluntary: Firm initiated