Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek...
Nidek Inc began a recall of Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek Multicolor Laser photocoagulator Model MC-500 is... on . FDA classified it as Class II and gives the reason as "When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0224-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nidek Incorporated
- Product
- Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek Multicolor Laser photocoagulator Model MC-500 is indicated for use in retinal photocoagulation for treatment of ocular fundus diseases.
- Product code
- HQF Laser, Ophthalmic
- Reason for recall
- When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Nidek Field Service engineers visited each affected location to upgrade software.
- Quantity in commerce
- 21affected devices
- Distribution
- US Distribution in the states of: NJ, SC, IL, MI, NY, CA, TX, OH, IN, UT, and MN.
- Product codes and lots
- Model MC-500 Serial numbers: 50277 50310 50330 50331 50357 50358 50383 50408 50417 50418 50425 50426 50439 50440 50445 50446 50471 50514 50311 50359 50382,
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0224-2015 | Class II | Terminated | Voluntary: Firm initiated |