Rapid Neg Urine Combo Panel Type 1; Catalog number B 1017-167; Siemens Material Number 10444652 MicroScan...
Siemens Healthcare Diagnostics, Inc. began a recall of Rapid Neg Urine Combo Panel Type 1; Catalog number B 1017-167; Siemens Material Number 10444652 MicroScan Rapid panels are designed for use in determining... on . FDA classified it as Class I and gives the reason as "An increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Neg Urine Combo Panel Type 1 used with MicroScan Microbiology Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0171-2015
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Rapid Neg Urine Combo Panel Type 1; Catalog number B 1017-167; Siemens Material Number 10444652 MicroScan Rapid panels are designed for use in determining identification to the species level and/or antimicrobial agent susceptibility of rapidly growing aerobic and facultatively anaerobic gram negative bacilli.
- Product code
- LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
- Reason for recall
- An increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Neg Urine Combo Panel Type 1 used with MicroScan Microbiology Systems.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Customer Recall notification letters were sent on October 17, 2014. Letters recommend that any remaining inventory be discarded and that consideration be given to reviewing previous test result, conduct patient follow up and/or repeat testing if isolates are still available. A second letter clarifying the recall was issued November 7, 2014 which provided additional information as requested by CDRH.
- Quantity in commerce
- 2239
- Distribution
- Distributed US (nationwide) including the states of IA, CA, NY, WI, MI, SD, NC, FL, HI, OH, WA, TX, WA, IN, NM, MS, KY, IL, UT, ND, and WY and the country of Canada.
- Product codes and lots
- Catalog number: B1017-167, Siemens Material number 1044652, lot numbers: 2014-11-01, Exp. 2014-11-01; 2015-01-03, Exp 2015-01-03; 2015-02-14, Exp 2015-02-14; 2015-03-21, Exp 2015-03-21; 2015-05-20, Exp 2015-05-20; 2015-06-18, Exp 2015-06-18; 2015-07-18, Exp 2015-07-18; 2015-08-11, Exp 2015-08-11/
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0170-2015 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0171-2015 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.