Z-0171-2015Class I

Rapid Neg Urine Combo Panel Type 1; Catalog number B 1017-167; Siemens Material Number 10444652 MicroScan...

Siemens Healthcare Diagnostics Inc. / initiated 2014-10-17 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Rapid Neg Urine Combo Panel Type 1; Catalog number B 1017-167; Siemens Material Number 10444652 MicroScan Rapid panels are designed for use in determining... on . FDA classified it as Class I and gives the reason as "An increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Neg Urine Combo Panel Type 1 used with MicroScan Microbiology Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0171-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Rapid Neg Urine Combo Panel Type 1; Catalog number B 1017-167; Siemens Material Number 10444652 MicroScan Rapid panels are designed for use in determining identification to the species level and/or antimicrobial agent susceptibility of rapidly growing aerobic and facultatively anaerobic gram negative bacilli.
Product code
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason for recall
An increase in false positive susceptible results on Rapid Neg BP Combo Panel Type 3 and Rapid Neg Urine Combo Panel Type 1 used with MicroScan Microbiology Systems.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Customer Recall notification letters were sent on October 17, 2014. Letters recommend that any remaining inventory be discarded and that consideration be given to reviewing previous test result, conduct patient follow up and/or repeat testing if isolates are still available. A second letter clarifying the recall was issued November 7, 2014 which provided additional information as requested by CDRH.
Quantity in commerce
2239
Distribution
Distributed US (nationwide) including the states of IA, CA, NY, WI, MI, SD, NC, FL, HI, OH, WA, TX, WA, IN, NM, MS, KY, IL, UT, ND, and WY and the country of Canada.
Product codes and lots
Catalog number: B1017-167, Siemens Material number 1044652, lot numbers: 2014-11-01, Exp. 2014-11-01; 2015-01-03, Exp 2015-01-03; 2015-02-14, Exp 2015-02-14; 2015-03-21, Exp 2015-03-21; 2015-05-20, Exp 2015-05-20; 2015-06-18, Exp 2015-06-18; 2015-07-18, Exp 2015-07-18; 2015-08-11, Exp 2015-08-11/