Integra Selector Kit Sterile single use device A bi-lumen, sterile single use accessory used for the Selector...
Integra LifeSciences Corp. began a recall of Integra Selector Kit Sterile single use device A bi-lumen, sterile single use accessory used for the Selector Integra Ultrasonic Surgical Aspirator system. One... on . FDA classified it as Class II and gives the reason as "Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0491-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Integra Selector Kit Sterile single use device A bi-lumen, sterile single use accessory used for the Selector Integra Ultrasonic Surgical Aspirator system. One of the tubing lumen transmits irrigation fluid from the console to the handpiece of the system, and the other tubing lumen transmits aspirated fluid and tissue back to the system console. The aspirator is intended to be used to facilitate the removal of cellular and other unwanted soft tissue. The system provides selective tissue disintegration with simultaneous irrigation and aspiration. The system has been designed for use by surgeons in the areas of neurosurgery and gastroenterology including laparoscopic procedures. Its use in other procedures is regarded as experimental and is thus subjecto to appropriate local regulatory approval
- Product code
- LFL Instrument, Ultrasonic Surgical
- Reason for recall
- Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Integra sent an 'Urgent- Medical Device Correction' letters/acknowledgement forms dated 9/29/2014 by traceable courier service. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory for the affected product, complete the attached form. Complete the other information on the form and return the form by email or fax. Customers with questions should contact Customer Service at 1-855-532-1723.
- Quantity in commerce
- 6,358 tubing kits
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Australia, Bulgaria, Canada, Germany, Great Britain, Greece, Hungary, Indonesia, Israel, Italy, Japan, Mexico, Portugal, Spain, Sweden, Thailand, Turkey, and Wales.
- Product codes and lots
- Lot Numbers: 130849;1132912;1132913;1132914;1133101;1133102;1133103;1133104;1133105;1133495;1134005;1134006;1134007;1134120;1134121;1134122;1134124;1134126;1134128;1134130;1134131;1134132;1134133;1134620;1134621;1134669;1134670;1134671;1134671;1134694;1134695;1140357;1141561;1141562;1141645;1141754;1141563;1141564;1141647;1142352;1142353;1142409; & 1142410
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0491-2015 | Class II | Terminated | Voluntary: Firm initiated |