Z-0463-2015Class II

Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing

Abbott Molecular, Inc. / initiated 2014-10-23 / Terminated / root cause: software

Abbott Molecular began a recall of Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing on . FDA classified it as Class II and gives the reason as "Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0463-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Molecular, Inc.
Product
Abbott m2000sp, an automated system for performing sample preparation for nucleic acid testing.
Product code
JJH Clinical Sample Concentrator
Reason for recall
Some versions of Application Specifications (App Spec) are incompatible with m2000sp system software version 6.0 and 7.0. This issue will cause Error Code 9000 (An unexpected error has occurred: Software error) to be generated when the Available Reagent Packs screen is selected.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
Field Correction Recall / Urgent Field Safety Notice letters dated September 23, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to: 1) review the app spec files that are installed on the m2000sp system and delete any incompatible app specs; 2) if one or more app spec files were deleted, installation of new app spec file(s) is needed; and, 3) review this information with laboratory personnel and retain the letter for future reference. Customers with questions about this market action can contact Technical Support at 800-553-7042, Option 2, or the local Abbott Molecular representative.
Quantity in commerce
9 CD ROMs (US); 1,273 CD ROMs (Foreign)
Distribution
Worldwide distribution. US nationwide (CA, CT, FL, NC, PA, SC, and DC), Algeria , Australia , Austria , Belgium , Bosnia and Herzegovina , Cameroon , Canada , Chad , Colombia , Costa Rica , Croatia , Denmark , Egypt , Ethiopia , Finland , France , Germany , Greece , Guinea , Honduras , Hong Kong , India , Indonesia , Ireland , Israel , Italy , Japan , Kenya , Korea , Lithuania , Mali , Mexico , Netherlands , New Zealand , Nigeria , Norway , Pakistan , Paraguay , Poland , Portugal , Romania , Russia , Saudi Arabia , Senegal , Slovenia , South Africa , Spain , Sweden , Switzerland , Taiwan , Tanzania , Thailand , Turkey , Uganda , Ukraine , United Arab Emirates, United Kingdom , Vietnam
Product codes and lots
List Number: 09K14; Models: E series, G series

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0463-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0463-2015Class IITerminatedVoluntary: Firm initiated