Z-0232-2015Class III
Etest Ceftaroline (CPT32) US B30. Etest (Antimicrobial Susceptibility Testing) is a quantitative technique...
bioMerieux, Inc. / initiated 2014-10-16 / Terminated / root cause: labeling and UDI
Biomerieux began a recall of Etest Ceftaroline (CPT32) US B30. Etest (Antimicrobial Susceptibility Testing) is a quantitative technique for determining the antimicrobial susceptibility of... on . FDA classified it as Class III and gives the reason as "The products have a wrong expiration date on their labeling, 5 years instead of 2 years". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0232-2015
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- Etest Ceftaroline (CPT32) US B30. Etest (Antimicrobial Susceptibility Testing) is a quantitative technique for determining the antimicrobial susceptibility of Gram negative and Gram positive aerobic bacteria.
- Product code
- JWY Manual Antimicrobial Susceptibility Test Systems
- Reason for recall
- The products have a wrong expiration date on their labeling, 5 years instead of 2 years.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- Consignees of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on October 16th, 2014. Additionally, the bioMrieux United States subsidiary notified US consignees of the FSCA notification via FedEx on October 23rd, 2014.
- Quantity in commerce
- 652 devices
- Distribution
- Worldwide Distribution-US (nationwide) including DC and the states of AL, AK, AR, AZ, CA, CO, DE, FL, GA, IA, IN, KY, MA, MD, MI, MO, NC, NE, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and the countries of Australia, Austria, China, Czech Republic, France, Germany, India, Italy, Japan, Netherlands, Poland, Portugal, Switzerland, Sweden, Turkey, United Kingdom, Egypt, Pakistan, Slovenia and Uruguay.
- Product codes and lots
- Ref #537540, Lot #1003202200, Right Expiry date: 27-APR-2016
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0230-2015 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0231-2015 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0232-2015 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0233-2015 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0234-2015 | Class III | Terminated | Voluntary: Firm initiated |