COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological
Iba Dosimetry Gmbh began a recall of COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological on . FDA classified it as Class II and gives the reason as "Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0168-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- IBA Dosimetry GmbH
- Product
- COMPASS, Model No. SW Version 3.1, catalogue number CS10-100 (medical Linear accelerator) radiological
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Error in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The field safety notice including a confirmation of receipt was sent to the affected hospitals; by email and UPS letter on 09/25/2014.
- Quantity in commerce
- 124 units
- Distribution
- Worldwide Distribution-Distributed in the US including Guam and the countries of Germany, Austria, France, UK, Poland, Finland, The Netherlands, Spain, Turkey, Russia, and South Africa, Algeria, Mexico, Brazil, Venezuela, Columbia, Panama, China, Japan Philippines, Korea, India, Singapore, Thailand, Taiwan, Hong Kong, and Malaysia.
- Product codes and lots
- N/A
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0168-2015 | Class II | Terminated | Voluntary: Firm initiated |