Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0582-2015Class II

Align Radial Stems, Prosthesis, Elbow, Hemi Replacement of the radial head for degenerative or post-traumatic...

Skeletal Dynamics, Inc / initiated 2014-10-16 / Terminated / root cause: design

Skeletal Dynamics began a recall of Align Radial Stems, Prosthesis, Elbow, Hemi Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and... on . FDA classified it as Class II and gives the reason as "Report received where the Align Radial Stem fractured". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0582-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Skeletal Dynamics, Inc
Product
Align Radial Stems, Prosthesis, Elbow, Hemi Replacement of the radial head for degenerative or post-traumatic disabilities presenting pain, crepitation, and decreased motion at the radio-humeral and/or proximal radio-ulnar joint with: Joint destruction and/or subluxation and resistance to conservative treatment. Intended use is also for primary replacement after fracture of the radial head and symptomatic sequelae after radial head resection.
Product code
KWI Prosthesis, Elbow, Hemi-, Radial, Polymer
Reason for recall
Report received where the Align Radial Stem fractured.
Root cause tag
design
FDA root cause
Device Design
Action
Skeletal Dynamic sent an Urgent Medical Device Safety Alert dated October 16, 2014, to all affected customers via certified mail. The Safety Alert included the warning "that radial head prostheses cannot be expected to withstand the activity levels and loads of normal healthy bone and joint tissue. Failure of the component can occur as a result of loss of fixation, strenuous activity, malalignment, trauma, non-union or excessive loads (estimated body weight equivalent of 350 lbs or greater)" with instructions to notify any surgeons that have used the product. A response form has been included with instructions to return the signed form to acknowledge the receipt and understanding of the new warning. Product labeling has been revised to include these labels in future shipments. Customers with questions were instructed to contact Customer Service at 877-753-5396 for assistance. For questions regarding this recall call 305-596-7585, ext 7026.
Quantity in commerce
1700
Distribution
Nationwide Distribution including GA, NJ, NM, IL, CA, OK, CO, NY, LA, OH, FL, VA, MD, MS, TX, IA, AL, WA, TN, ME, CT, WI, AZ, NE, SC, NC and Puerto Rico.
Product codes and lots
Cat #s: ALN-RST-0700, ALN-RST-0702, ALN-RST-0704, ALN-RST-0706, ALN-RST-0708, ALN-RST-0800, ALN-RST-0802, ALN-RST-0802, ALN-RST-0804, ALN-RST-0806, ALN-RST-0808, ALN-RST-0900, ALN-RST-0902, ALN-RST-0904, ALN-RST-0906, ALN-RST-0908, ALN-RST-1000, ALN-RST-1002,  ALN-RST-1004, ALN-RST-1006, ALN-RST-1008, ALN-RST-1100, ALN-RST-1102, ALN-RST-1104, ALN-RST-1106, ALN-RST-1108; Lot #s: AN0610023, AN0212065, AN0913033, AN0913034, AN0610024, AN0913035, AN0913036, AN0610025, AN0913037, AN0913038, AN0610026,  AN0913039, AN0913040, AN0610027, AN0913041, AN0913042, AN0610028, AN0212066, AN0913043,  AN0913044, AN0610029, AN0913045, AN0913046,  AN0610030, AN0913047, AN0913048, AN0610031,  AN0913049, AN0913050, AN0610032, AN0913051,  AN0913052, AN0610033, AN0212067, AN0913053, AN0913054, AN0610034, AN0913055, AN0913056, AN0610035, AN0913057, AN0913058, AN0610036,  AN0913059, AN0913060, AN0610037, AN0913061,  AN0913062, AN0610038, AN0212068, AN0913063, AN0913064, AN0610039, AN0212069, AN0913065, AN0913066, AN0610040, AN0913067, AN0913068,  AN0610041, AN0913069, AN0913070, AN0410008,  AN0610042, AN0410009, AN0913071, AN0913072,  AN0610043, AN0212070, AN0913073, AN0913074,  AN0610044, AN0913075, AN0913076, AN0610045, AN0913077, AN0913078, AN0610046, AN0913079, AN0913080, AN0610047, AN0913081, AN0913082.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0582-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0582-2015Class IITerminatedVoluntary: Firm initiated