Z-0932-2015Class II
Baxter Amia Automated PD systems are used in the treatment of adult renal failure patients to perform...
Baxter Healthcare Corporation / initiated 2014-04-24 / Terminated
Baxter Healthcare Corp. began a recall of Baxter Amia Automated PD systems are used in the treatment of adult renal failure patients to perform automated peritoneal dialysis on . FDA classified it as Class II and gives the reason as "System error 01779 is produced when the battery cannot be charged due to specific voltage differences between the battery and the charger". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0932-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Baxter Amia Automated PD systems are used in the treatment of adult renal failure patients to perform automated peritoneal dialysis.
- Product code
- FKX System, Peritoneal, Automatic Delivery
- Reason for recall
- System error 01779 is produced when the battery cannot be charged due to specific voltage differences between the battery and the charger.
- FDA root cause
- Under Investigation by firm
- Action
- Baxter sent an IMPORTANT PRODUCT INFORMATION letter dated April 24, 2014 to all affected customers. The peritoneal dialysis provider consignees were instructed to: 1) complete the enclosed customer reply form and return it to Baxter; and, 2) contact Baxter Technical Services at 800-553-6898 should a System Event 01179 occur. Baxter Technical Services is available 24 hours a day, 7 days a week. Clinical questions can be directed to Baxter's Renal Clinical Helpline at 888-736-2543, option 2 (Monday - Friday, 8:00 AM - 5:00 PM CT). General questions regarding the Import Product Information letter can be direct to The Center for One Baxter at 800-422-9837 (Monday - Friday, 8:00 AM - 5:00 PM CT). For questions regarding this recall call 800-422-9837.
- Quantity in commerce
- 55 units
- Distribution
- Nationwide Distribution including GA, MA, NC, NY, PA, and WA.
- Product codes and lots
- Product Code: 5C9310
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0932-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.