Z-0958-2015Class II
Philips Expression MR200 MRI Patient Monitoring System
Invivo Corporation / initiated 2014-10-22 / Terminated / root cause: design
Invivo Corporation began a recall of Philips Expression MR200 MRI Patient Monitoring System on . FDA classified it as Class II and gives the reason as "The MR 200 devices failed the initial power up test due to capacitor short and hi-pot testing; possible loss of Patient Monitoring or User/Service Electrical Shock". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0958-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Invivo Corporation
- Product
- Philips Expression MR200 MRI Patient Monitoring System.
- Product code
- MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
- Reason for recall
- The MR 200 devices failed the initial power up test due to capacitor short and hi-pot testing; possible loss of Patient Monitoring or User/Service Electrical Shock.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Philips Medical Systems sent an Urgent - Medical Device Correction Letter dated October 22, 2014, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. Our records indicate that you have an affected device. The following page provides additional instructions and actions that will be taken to address this problem. If you need any further information or support concerning this issue, please contact your local Philips representative: 877-468-4861 option 1, then option 2.
- Quantity in commerce
- 20
- Distribution
- CA, FL, IN, KY, MN, NC, NJ, NY, OK, SD and TX Netherlands, Poland and United Kingdom.
- Product codes and lots
- Model No. 866120; Lot/ Serial Numbers: US33600065, US33600066, US33600075, US33600077, US33600078, US33600079; US33600080, US33600081, US33600082, US33600083, US33600084, US33600086, US33600087, US33600088, US33600089, US33600090, US33600091, US33600092, US33600093, US33600094.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0958-2015 | Class II | Terminated | Voluntary: Firm initiated |