510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0226-2015Class I

Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction...

Synthes (USA) Products LLC / initiated 2014-10-23 / Terminated / root cause: manufacturing process

Synthes (USA) Products LLC began a recall of Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System... on . FDA classified it as Class I and gives the reason as "Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0226-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes (USA) Products LLC
Product
Synthes Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws: The Titanium Polyaxial Reduction Head is a part of the Matrix Spine System. Intended for posterior pedicle screw fixation (Tl OS2- ilium), posterior hook fixation (Tl-L5), or anterolateral fixation (T8-L5).
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not have been machined to the correct specification.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
DePuy Synthes sent an Urgent Notice: Medical Device Recall letter, dated October 23, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check inventory and remove affected stock, call Synthes to obtain a Return Authorization for the device, and to return the completed verification section of the response form by fax to (610) 430-7083 or Scan/email: [email protected]. Customers with questions were instructed to call 610-719-5450.
Quantity in commerce
24
Distribution
Distributed to MI, MT, and CO.
Product codes and lots
Part Number  04.634.002   Lot Number 6816781

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0226-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0226-2015Class ITerminatedVoluntary: Firm initiated